4 Troubleshooting
HB-005615-j
70
Problem
Possible Cause(s)
Recommended action(s)
suspected
(see Status
Code DP02 in
memory
dump)
Gasket of Docking
Station
contaminated
Clean the Docking Station gasket with a
cotton swab with 70% isopropanol. Remove
any residual threads and fibers and allow to
dry.
Note: Always clean the sensor before
inserting it into the Docking Station.
Gasket of Docking
Station defective or
missing
Carefully inspect the gasket and verify that it
is present and well embedded in its notch.
If damaged contact qualified service
personnel.
Docking Station
door and / or
closing mechanism
defective
Carefully inspect the door and the closing
mechanism for mechanical or functional
damage. The closing mechanism must
properly lock in place.
If damaged contact qualified service
personnel.
Docking Station
door mechanically
defective
Replace Docking Station door according to
section
Gasket defective /
missing
Place gasket correctly in the chamber in the
Docking Station according to
section
“Replacement, Docking Station Gasket”.
SDM defective
Contact SenTec authorized service personnel.
Calibration Module
defective
Replace Calibration Module according to the
Repair Manual for the SDMS.
To be performed by SenTec authorized
service technicians only
.
Docking Station
Module defective
Replace Docking Station Module according to
the
Repair Manual for the SDMS.
To be performed by SenTec authorized
service technicians only
.
Note: Also refer to messages ‘Gas leak in DS’, ‘Leak test in
progress’ described in the
Technical Manual for the SDM.
P0503
Service Gas
bottle cannot
be installed
properly
Plastic lid of the
new Service Gas
bottle not removed
Remove the plastic lid and install the Service
Gas bottle.
Foreign body in the
slot for the Service
Gas bottle
Switch off the SDM, disconnect it from the
mains supply and remove the foreign body
from the slot on the rear panel of the SDM.