3 Maintenance/Safety & Functionality Tests
SenTec Digital Monitoring System
25
Slope Test Control Table PO2 – in mmHg
(after 6% and 21% O2 calibration)
Pcal = value after calibration with 12% O2
Table is only valid for sensor temp. of 42°C.
P
atm
Pcal 6_min
6_ideal
6_max
21_min
21_ideal
21_max
600
72.0
34
36
38
118
126
134
605
72.6
34
36
38
119
127
135
610
73.2
35
37
38
120
128
136
615
73.8
35
37
39
121
129
137
620
74.4
35
37
39
122
130
138
625
75.0
36
38
39
123
131
139
630
75.6
36
38
40
124
132
140
635
76.2
36
38
40
125
133
141
640
76.8
36
38
40
126
134
142
645
77.4
37
39
41
127
135
144
650
78.0
37
39
41
128
137
145
655
78.6
37
39
41
129
138
146
660
79.2
38
40
42
130
139
147
665
79.8
38
40
42
131
140
148
670
80.4
38
40
42
132
141
149
675
81.0
38
41
43
133
142
150
680
81.6
39
41
43
134
143
151
685
82.2
39
41
43
135
144
152
690
82.8
39
41
43
136
145
154
695
83.4
40
42
44
137
146
155
700
84.0
40
42
44
138
147
156
705
84.6
40
42
44
139
148
157
710
85.2
40
43
45
140
149
158
715
85.8
41
43
45
141
150
159
720
86.4
41
43
45
142
151
160
725
87.0
41
44
46
143
152
161
730
87.6
42
44
46
144
153
162
735
88.2
42
44
46
145
154
164
740
88.8
42
44
47
146
155
165
745
89.4
42
45
47
147
156
166
750
90.0
43
45
47
148
158
167
755
90.6
43
45
48
149
159
168
760
91.2
43
46
48
150
160
169
765
91.8
44
46
48
151
161
170
770
92.4
44
46
49
152
162
171