Tools
24
4.2.6
Odometer Calibration
This test ensures that the odometer is measuring distance accurately. Even though the
odometer is calibrated in the factory, you must calibrate it periodically to ensure accuracy. This
is particularly important as the calibration can change with different surface materials.
When selecting this option, you will see the screen in Figure 4-16. The following options exist:
Figure 4-16: Screen displaying current odometer calibration
•
Odometer Test
– This test verifies the odometer is calibrated properly. It involves
moving the system a set dista
nce (5m or 10’, depending on units used), and comparing
the value obtained to a known value. The screen in Figure 4-17 will be displayed once
Odometer Test is pressed. Press
Start
to begin and then press
Finish
when you’ve
moved that distance. It will display a pass or fail, along with a percentage difference
from the known value (Figure 4-18).
Summary of Contents for LMX100
Page 1: ...2016 00048 05 ...
Page 2: ...ii ...
Page 4: ...Product Registration iv ...
Page 10: ...x ...
Page 14: ...Overview 2 ...
Page 40: ...Tools 28 ...
Page 44: ...GPR Concepts 32 ...
Page 58: ...Data Collection 46 ...
Page 63: ...Screenshots Mini Reports 51 Figure 7 6 Example of a Mini Report ...
Page 64: ...Screenshots Mini Reports 52 ...
Page 68: ...Transferring Data to a PC 56 ...
Page 72: ...Troubleshooting 60 ...
Page 75: ...Care Maintenance 63 ...
Page 76: ...Care Maintenance 64 ...
Page 78: ...Parts List Accessories 66 ...
Page 80: ...Parts List Accessories 68 ...
Page 82: ...Technical Specifications 70 ...
Page 84: ...Appendix A Health Safety Certification 72 ...
Page 92: ...Appendix B GPR Emissions Interference and Regulations 80 ...
Page 94: ...Appendix C Instrument Interference 82 ...
Page 96: ...Appendix D Safety around Explosive Devices 84 ...