MT4299E Rev.1
15/10/2019
Pagina 7 di 55
SMART P-2 detector for Carbon Monoxide and Nitrogen Dioxide/Petrol
Vapours
CODE
DESCRIPTION
CAT.
€
S2400CO-ND
CARBON MONOXIDE & NITROGEN DIOXIDE detector for car
parks,
RS485 serial output.
Range 0-300 ppm CO and 0-30
ppm NO2.
B
400,00
S2400CO-VB
CARBON MONOXIDE & PETROL VAPOURS detector for car
parks,
RS485 serial output.
Range 0-300 ppm CO and 0-
100%LEL Petrol Vapours.
B
420,00
CALIBRATION TOOLS for SMART P
CODE
DESCRIPTION
CAT.
€
STS/CKD-PK
Handheld calibration keypad to be connected to the detector to
adjust the Zero, Span and 4-20mA values.
A
200,00
2.1)
Technical specifications
Typical use
To be used with MU+PK
256/128/64/32 type units to monitor toxic gas in
car parks and garages
Detectable gas
Carbon Monoxide 0-300 ppm.
Nitric dioxide 0 - 30 ppm
Petrol Vapours 0-100% LEL
Sensor used
Electrochemical cell for CO and NO2. Catalytic
for Petrol Vapour.
Power
12-24 Vdc (-20% +15%)
Proportional output
RS485 serial
Absorption at
12-24Vdc
CO max 70mA
NO2 max 70mA
CO+NO2 max 80mA
CO+PETROL VAPOUR max 110 mA
Repeatability
5% FS
Storage temp.
-25 / + 60 °C
Working temperature
-20 / + 60 °C
Relative humidity
20-90 % non condensed RH/40°C
IP protection level
IP55
Weight
600 gr.
Dimensions
L. 100, H. 180, D. 65 mm
Estimated sensor life
4 years for CO and PETROL VAPOUR sensors
3 years for NO2 sensor