Sensit SCAL-400 Instruction Manual Download Page 5

5

4 / 12 

Description 

The

 

SCal

400

 

is

 

a

 

multi

unit

 

charging

 

and

 

calibration

 

station

 

for

 

Sensit

 

P400

 

gas

 

detection

 

instruments.

 

It

 

can

 

bump

 

test

 

and/or

 

calibrate

 

up

 

to

 

five

 

P400

 

units

 

consecutively

 

while

 

simultaneously

 

recharging

 

the

 

battery

 

packs.

 

The

 

SCal

400

 

also

 

detects

 

if

 

a

 

P400

 

requires

 

calibration

 

or

 

a

 

bump

 

test

 

when

 

connected

 

and

 

automatically

 

performs

 

the

 

required

 

test.

   

During

 

connection

 

any

 

manually

 

performed

 

calibration

 

and

 

bump

 

tests

 

in

 

the

 

instrument

 

memory

 

log

 

will

 

automatically

 

transfer

 

from

 

P400

 

to

 

SCal

400.

 

Data

 

management

 

for

 

the

 

SCal

400

 

requires

 

either

 

SCal

D

 

or

 

SCal

N

 

software

 

packages.

   

All

 

SCal

400

 

stations

 

come

 

standard

 

with

 

SCal

D

 

software

 

for

 

desktop

 

application.

 

The

 

location

 

to

 

place

 

any

 

instrument

 

is

 

referred

 

to

 

as

 

a

 

“slot”.

   

For

 

each

 

slot

 

there

 

are

 

charger

 

and

 

communication

 

connections

 

as

 

well

 

as

 

LEDs

 

to

 

indicate

 

the

 

status.

   

All

 

testing

 

functions

 

require

 

no

 

user

 

interaction.

 

The

 

display

 

is

 

a

 

touchscreen.

   

Instrument

 

and

 

station

 

information

 

can

 

be

 

accessed

 

by

 

using

 

your

 

finger

 

or

 

the

 

included

 

stylus.

 

The

 

SCal

400

 

scans

 

each

 

slot,

 

updates

 

the

 

LED

 

status,

 

and

 

refreshes

 

the

 

display

 

screen

 

for

 

each

 

slot

 

every

 

three

 

seconds.

 

During

 

communication,

 

the

 

SCal

400

 

completes

 

any

 

log

 

transfer

 

before

 

scanning

 

another

 

slot.

   

 

 

LED Status 

Each

 

slot

 

has

 

two

 

LEDs

 

indicating

 

the

 

status

 

for

 

the

 

docked

 

instrument.

 

The

 

following

 

table

 

describes

 

the

 

various

 

LED

 

states:

   

SCal

400

 

Individual

 

Slot

 

LED

 

Status:

   

Green

 

LED

 

Red

 

LED

 

Description

 

1

 

OFF

 

OFF

 

Nothing

 

on

 

slot

 

2

 

ON

 

OFF

 

Ready

 

(all

 

test

 

done,

 

fully

 

charged/alkaline

 

OK)

 

3

 

OFF

 

ON

 

Fail

 

calibration/bump

 

test/alkaline

 

low

 

battery

 

4

 

BLINK

 

ON/OFF/BLINK

 

Charging

 

5

 

ON/BLINK

 

BLINK

 

Test

 

in

 

progress/pending

 

 

 

 

Summary of Contents for SCAL-400

Page 1: ...N D E R S T A N D I N S T R U C T I O N S B E F O R E U S E FOR USE IN AREAS KNOWN TO BE FREE OF COMBUSTIBLE GAS HAZARDS FOR USE IN AREAS THAT ARE CLEAN AND DRY 851 Transport Drive Valparaiso IN 4638...

Page 2: ......

Page 3: ...en 7 Main Menu Screen 8 Calibration Gas Information Screen 8 Log Review Screen 9 Password Screen 9 Supervisor Menu Screen 10 12 Operation Set up 13 Operation System Activation 13 Operation P400 Testin...

Page 4: ...emperature 40 122 F 5 50 C Maximum inlet pressure 20psi Features Touch screen display LED status indicators Charging pins Communication pin Power connection RJ45 Serial Connections Cooling fan intake...

Page 5: ...t is referred to as a slot For each slot there are charger and communication connections as well as LEDs to indicate the status All testing functions require no user interaction The display is a touch...

Page 6: ...Screen only if no unit test is pending or in progress Sample image of the Main Screen Table 1 Main Screen background colors and descriptions Slot Screen Background color Status Description GRAY Empty...

Page 7: ...ttery WHITE CHARGING RED ALKALINE LOW BATTERY GREEN CHARGED ALKALINE OK 3 Data Transfer GRAY PENDING WHITE UPDATING RTC DOWNLOADING CALIBRATION LOG DOWNLOADING BUMP TEST LOG GREEN COMPLETED DATA TRANS...

Page 8: ...eight cells Each cell represents a single gas inlet valve The background colors and gas inlets are color coded and matched for easy identification The colors on the manifold label correspond to the co...

Page 9: ...omatically be checked Upon providing the correct password the next screen will be displayed Entering an incorrect password will change the screen title to be displayed as WRONG PASSWORD in red After t...

Page 10: ...SCal 400 application All available time zones will be listed Select the appropriate time zone and click UPDATE A message box will be displayed to indicate system will restart Click OK to continue 2 S...

Page 11: ...be present A copy of the database will be saved to the memory card every day The automatically saved backup files are named AUTO_DDMMYYYY sdf b BACKUP DATABASE This option will create a manual backup...

Page 12: ...libration bump test 12 PURGE TIME AFTER Amount of time in seconds the unit will purge the line with air after it completes a calibration bump test 13 SPECIALITY GAS PURGE Amount of time in seconds the...

Page 13: ...ot 4 If the instrument is off it will turn on automatically 5 The instrument will be bump tested and or calibrated as determined by the instrument setup 6 The instrument will recharge if needed Alkali...

Page 14: ...NOTES...

Page 15: ...NOTES...

Page 16: ...before warranty is rendered Units out of warranty will be repaired for a service charge Internal repair or maintenance must be performed by a Sensit Technologies authorized technician Violation will...

Reviews: