ADJUSTABLE FEATURES
The Above settings are field adjustable:
Setting Display Mode For %LEL & Alarm Setpoint For %LEL
1.
While the unit is on and in the working display, push the A and C buttons until the
display shows OPt.
2.
Press the B button once until the display shows diS%. This indicates the working
display will show measurements in %LEL.
3.
Press the B button once more until the display shows AL%. This indicates the
alarm setpoint can be set in units of %LEL.
4.
Press the B button once more until the display shows XX.X%. This is the alarm
setpoint (XX.X can be values of 0-9).
5.
Press the C button to increment the alarm setpoint or press the B button to
decrement the alarm setpoint (Minimum 2.0%LEL, Maximum 60.0%LEL)
6.
Press the A button once to exit to the OPt menu.
7.
Press the A button once more to exit to the working display which will show
measurements in %LEL.
Setting Display Mode For PPM & %LEL & Alarm Setpoint For PPM
1.
While the unit is on and in the working display, push the A and C buttons until the
display shows OPt.
2.
Press the B button once until the display shows diS%. This indicates the working
display will show %LEL measurements.
3.
Press the C button once until the display show diS% ppm. This indicates the
working display will initially show ppm measurements then transition to %LEL
measurements once the unit measures greater than 990ppm.
4.
Press the B button once more until the display shows AL ppm. This indicates the
alarm setpoint can be set in units of PPM.
5.
Press the B button once more until the display shows XXXppm. This is the alarm
setpoint; it must be set to a PPM value greater than or equal to 100ppm to
activate the PPM alarm.
6.
Press the C button to increment the alarm setpoint or press the B button to
decrement the alarm setpoint (PPM Alarm Off: 99ppm, Minimum 100ppm,
Maximum 990ppm)
7.
Press the A button once to exit to the OPt menu.
8.
Press the A button once more to exit to the working display which will show
measurements in PPM.
SENSOR REPLACEMENT
Step 1: Turn off instrument.
Step 2: Remove sensor cap and sensor retainer filter.
Step 3: Remove sensor. Observe the location of the tab relative
to the white dot beneath, on the circuit board.
Step 4: Replace sensor. Be sure to align the sensor tab directly
over the white dot.
Step 5: Place sensor retainer filter inside sensor cap and install
sensor cap.
Step 6: Turn instrument on and wait for zero process to complete.
“bAd” indicates improperly installed sensor or additional warm-up
time is required.
Allow the instrument to operate for 10 minutes,
then press and hold the C button to reestablish zero. If “bAd”
remains on the display, remove the instrument from service.
Step 7:
If zeroing of the replacement sensor was successful
and at least a 5 minute warm-up has been observed, proceed
with the calibration instructions.
WARNING: Perform service only in areas known to be non-
flammable.
WARNING: Substitution of components may impair intrinsic safety.
WARNING: Use only sensor from SENSIT Technologies.
Part Number 375-2611-01
Alarm PPM
100-990
OFF
Alarm-LEL
2-60%
10%
Show PPM
10PPM Resolution
ON
MENU ITEMS
RANGE
DEFAULT