background image

12

13

CALIBRATION 

continued from page 12

Step 3: Attach the adapter to the sensor cap assembly.

Step 4: Press the TICK/CAL button (B) and the cal icon will flash
and numbers will change on the display. Upon completion, the
working display will appear with the gas reading.

A reading of “bAd” indicates unsuccessful calibration. Repeat
process if “bAd” appears. Press the POWER/MUTE button (A)
to clear the “bAd” display reading and go to the working display.

Continued unsuccessful calibrations may indicate faulty sensor.

Any instrument that will not calibrate or continues to indicate
“bAd” on the display, should be taken out of service. Please
contact Sensit Technologies for further assistance.

CALIBRATION CHECK

To verify the accuracy of any Sensit® HXG-2d it must be ex-
posed to a known concentration of test gas.  Any sensor that
does not meet the specifications listed in this manual may re-
quire calibration or replacement.

Any time it is suspected the Sensit® HXG-2d is not working properly,
has read “OL” or has been exposed to silicone, check calibration.

Calibration is the process of setting the readings of the instru-
ment to equal the value of the certified calibration gas. The in-
strument should operate for 5 minutes before calibrating.

NOTE: Use of calibration kits other than those offered or ap-
proved by SENSIT TECHNOLOGIES may cause inaccurate read-
ings. Repairs may be required if the instrument fails to calibrate.
Use only sensors supplied by SENSIT TECHNOLOGIES or their
authorized representative.

NOTE: During the calibration process the numbers on the dis-
play are not the actual gas concentration.

Step 1: Prepare 50%LEL or 10% LEL Methane/Air, depending
on what calibration point the unit calls for, and proper regulator
and adapter.

Step 2: Press the TICK/CAL button (B) for 6 seconds until  “GAS”
is displayed.

CALIBRATION

(Continued on page 13)

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for 906-00000-08

Page 1: ...inop erative from such defects the unit will be repaired or replaced at our option This warranty covers normal use and does not cover damage which occurs in shipment or failure which results from alte...

Page 2: ...ion which if not avoided may result in minor or moderate injury NOTICE WARNING This safety symbol is used to indicate a potentially hazardous situa tion which if not avoided may result in serious inju...

Page 3: ...of combustible gases in both the PPM and LEL range The PPM readings auto range to LEL when the concentration exceeds 990PPM 2 LEL meth ane Optionally all readings can be set to a resolution of 0 1 LEL...

Page 4: ...battery compartment All Sensit HXG 2d instruments require 3 Duracell PC1400 batteries provide approximately 50 hours of continuous use A push button located in the center of the instrument activates a...

Page 5: ...ly inoperable or improperly zeroed at start up the display will show bAd BATTERYINSTALLATION REPLACEMENT 9 Continued on page 10 OPERATIONAND USE WARNING Always change batteries in an environment free...

Page 6: ...ark areas press and release the LIGHT ZERO button C to turn on the back light Press again to turn off 12 To turn instrument off push and hold the POWER MUTE button A for 5 6 seconds until OFF appears...

Page 7: ...r that does not meet the specifications listed in this manual may re quire calibration or replacement AnytimeitissuspectedtheSensit HXG 2disnotworkingproperly has read OL or has been exposed to silico...

Page 8: ...etpoint can be set in units of PPM 5 Press the B button once more until the display shows XXXppm This is the alarm setpoint it must be set to a PPM value greater than or equal to 100ppm to activate th...

Reviews: