     CAUTION: Always start any Sensit® HXG-2d in a gas free

environment to ensure a proper zero.

1. Push and hold the power button (A) until the display and backlight
illuminates.

2. If the display fails to illuminate or the BAT icon is shown on the
display, replace the batteries.

3. During successful start-up the instrument will display:

a. Display all segments
b. Turn on and off the backlight
c. Display “S2d” as the model number
d. Display software version
e. Display “50.0%” or “10.0%” indicating the LEL Calibration point

for the instrument

f. Display “dIS” indicating display resolution of PPM / %LEL

or both.

g. Display “AL” followed by the alarm set point.
h. Activate the alarm sound and red LED for 3 beeps
i. Continue to flash all segments until proper warm-up is

attained (no more than five minutes)

j. Flash “Zro” indicating fresh air zeroing
k. Working display is shown

NOTE: A failure to properly zero due to the presence of gas is indicated
by “bAd” + cal icon illuminated on the display. Pressing the ZERO (C)
button for 6 seconds and releasing will restart the zeroing process.

NOTE: If a sensor is completely inoperable or improperly zeroed at
start up, the display will show “bAd”.

BATTERY INSTALLATION/REPLACEMENT

9

(Continued on page 10)

OPERATION AND USE

       WARNING: Always change batteries in an environment
free of combustible gases.

      WARNING: Do not mix batteries of different type or age.

Battery replacement is necessary when the BAT icon is illumi-
nated and the green LED is flashing. This warning indicates there
is approximately 15 minutes or less operational time before the
instrument turns off automatically due to insufficient battery power.

Remove the battery sleeve cover by removing the battery sleeve
screw with a T-10 screwdriver. Depress the locking tab on the front
of the handle with a coin or flat object and pull the battery sleeve
handle away from the top or display area of the instrument.

Place 3 approved batteries into the battery holder. Observe the
polarity markings on the inside of the battery holder for proper
battery installation. Improper installation will cause the instrument
not to operate. Replace the battery sleeve and allow the locking
tab to snap into position. Replace the battery sleeve retaining screw.
Check to be sure the handle is secure to the instrument body by
firmly pulling the handle away. The handle will remain in place if a
proper connection is made.

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for 906-00000-08

Page 1: ...inop erative from such defects the unit will be repaired or replaced at our option This warranty covers normal use and does not cover damage which occurs in shipment or failure which results from alte...

Page 2: ...ion which if not avoided may result in minor or moderate injury NOTICE WARNING This safety symbol is used to indicate a potentially hazardous situa tion which if not avoided may result in serious inju...

Page 3: ...of combustible gases in both the PPM and LEL range The PPM readings auto range to LEL when the concentration exceeds 990PPM 2 LEL meth ane Optionally all readings can be set to a resolution of 0 1 LEL...

Page 4: ...battery compartment All Sensit HXG 2d instruments require 3 Duracell PC1400 batteries provide approximately 50 hours of continuous use A push button located in the center of the instrument activates a...

Page 5: ...ly inoperable or improperly zeroed at start up the display will show bAd BATTERYINSTALLATION REPLACEMENT 9 Continued on page 10 OPERATIONAND USE WARNING Always change batteries in an environment free...

Page 6: ...ark areas press and release the LIGHT ZERO button C to turn on the back light Press again to turn off 12 To turn instrument off push and hold the POWER MUTE button A for 5 6 seconds until OFF appears...

Page 7: ...r that does not meet the specifications listed in this manual may re quire calibration or replacement AnytimeitissuspectedtheSensit HXG 2disnotworkingproperly has read OL or has been exposed to silico...

Page 8: ...etpoint can be set in units of PPM 5 Press the B button once more until the display shows XXXppm This is the alarm setpoint it must be set to a PPM value greater than or equal to 100ppm to activate th...

Reviews: