SENSIRION Winnipeg Ventilator Quick Reference Manual Download Page 5

Measure the (MALE) Pa�ent Height

Determine Tidal Volume using IBW Chart

See the User Manual Sec�on 3.2 for further detail

Ideal Body Weight

Male

4

Page 5 of 14

MAL

E

 – ID

E

AL 

BO

DY W

E

IGH

T

 A

N

T

IDAL V

O

L

U

M

E

 CHAR

T

 

PATIENT 

H

EIG

H

T

 

IBW 

 

FT’ IN” 

IN” 

CM 

KG 

4' 0" 

48 

122

 

22.4 

90 

112 

134 

157 

179 

4' 1" 

49 

124

 

24.7 

99 

124 

148 

173 

198 

4' 2" 

50 

127

 

27 

108 

135 

162 

189 

216 

4' 3" 

51 

130

 

29.3 

117 

147 

176 

205 

234 

4' 4" 

52 

132

 

31.6 

126 

158 

190 

221 

253 

4' 5" 

53 

135

 

33.9 

136 

170 

203 

237 

271 

4' 6" 

54 

137

 

36.2 

145 

181 

217 

253 

290 

4' 7" 

55 

140

 

38.5 

154 

193 

231 

270 

308 

4' 8" 

56 

142

 

40.8 

163 

204 

245 

286 

326 

4' 9" 

57 

145

 

43.1 

172 

216 

259 

302 

345 

4' 10"  58 

147

 

45.4 

182 

227 

272 

318 

363 

4' 11"  59 

150

 

47.7 

191 

239 

286 

334 

382 

5' 0" 

60 

152

 

50 

200 

250 

300 

350 

400 

5' 1" 

61 

155

 

52.3 

209 

262 

314 

366 

418 

5' 2" 

62 

157

 

54.6 

218 

273 

328 

382 

437 

5' 3" 

63 

160

 

56.9 

228 

285 

341 

398 

455 

5' 4" 

64 

163

 

59.2 

237 

296 

355 

414 

474 

5' 5" 

65 

165

 

61.5 

246 

308 

369 

431 

492 

5' 6" 

66 

168

 

63.8 

255 

319 

383 

447 

510 

5' 7" 

67 

170

 

66.1 

264 

331 

397 

463 

529 

5' 8" 

68 

173

 

68.4 

274 

342 

410 

479 

547 

5' 9" 

69 

175

 

70.7 

283 

354 

424 

495 

566 

5' 10"  70 

178

 

73 

292 

365 

438 

511 

584 

5' 11"  71 

180

 

75.3 

301 

377 

452 

527 

602 

6' 0" 

72 

183

 

77.6 

310 

388 

466 

543 

621 

6' 1" 

73 

185

 

79.9 

320 

400 

479 

559 

639 

6' 2" 

74 

188

 

82.2 

329 

411 

493 

575 

658 

6' 3" 

75 

191

 

84.5 

338 

423 

507 

592 

676 

6' 4" 

76 

193

 

86.8 

347 

434 

521 

608 

694 

6' 5" 

77 

196

 

89.1 

356 

446 

535 

624 

713 

6' 6" 

78 

198

 

91.4 

366 

457 

548 

640 

731 

6' 7" 

79 

201

 

93.7 

375 

469 

562 

656 

750 

6' 8" 

80 

203

 

96 

384 

480 

576 

672 

768 

6' 9" 

81 

206

 

98.3 

393 

492 

590 

688 

786 

6' 10"  82 

208

 

100.6 

402 

503 

604 

704 

805 

6' 11"  83 

211

 

102.9 

412 

515 

617 

720 

823 

7' 0" 

84 

213

 

105.2 

421 

526 

631 

736 

842 

GREY AREA INDICATES TARGET VOLUMES 

LESS THAN THE CAPABILITY 

OF THE VENTILATOR

6   

7   

TIDAL VOLUME (

mL/kg

) AT:

 

Summary of Contents for Winnipeg Ventilator

Page 1: ...Pressure Leak Test Ideal Body Weight MALE 5 Ideal Body Weight FEMALE Full IFU www CEVen lators ca Call Toll Free 1 833 229 1036 Circuit Setup 6 Inspiratory Time 7 I E 9 Alarm Troubleshoo ng Quick Refe...

Page 2: ...ways perform the Pre Use Pressure Test BEFORE connec ng a new circuit to a pa ent In PSV mode the expiratory trigger sensi vity is xed at 10 L min or 25 PEAK inspiratory ow Set the Low Inspiratory Pre...

Page 3: ...on the back of the ven lator To TURN OFF the ven lator place the ven lator in standby mode then push and hold the power bu on at the back of the ven lator for 3 seconds Pa ent Inspiratory Port with in...

Page 4: ...ltaneously for 2 seconds Verify that the Pressure graph reaches 50 cmH2 O See the Insp Tidal Volume display for test status and result If the system fails the pressure test check the system for leaks...

Page 5: ...2 62 157 54 6 218 273 328 382 437 5 3 63 160 56 9 228 285 341 398 455 5 4 64 163 59 2 237 296 355 414 474 5 5 65 165 61 5 246 308 369 431 492 5 6 66 168 63 8 255 319 383 447 510 5 7 67 170 66 1 264 3...

Page 6: ...2 5 2 62 157 50 1 200 251 301 351 401 5 3 63 160 52 4 210 262 314 367 419 5 4 64 163 54 7 219 274 328 383 438 5 5 65 165 57 228 285 342 399 456 5 6 66 168 59 3 237 297 356 415 474 5 7 67 170 61 6 246...

Page 7: ...0 73 0 66 0 60 0 55 0 51 0 47 0 44 0 41 0 39 0 37 0 35 0 33 0 31 0 30 0 29 0 28 600 1 80 1 44 1 20 1 03 0 90 0 80 0 72 0 65 0 60 0 55 0 51 0 48 0 45 0 42 0 40 0 38 0 36 0 34 0 33 0 31 0 30 650 1 95 1...

Page 8: ...Results in impossible se ng where respiratory period is less than inspiratory me 12 14 16 18 20 22 24 26 28 30 35 9 0 7 6 6 5 5 7 5 0 4 5 4 0 3 6 3 3 3 0 2 4 7 3 6 1 5 3 4 6 4 0 3 5 3 2 2 8 2 6 2 3 1...

Page 9: ...50 40 30 20 0 20 40 60 80 100 120 140 160 180 200 5 10 15 20 7 5 25 L min L min L min L min L min L min O 2 se ling me delay on the Winnipeg Ven lator is a func on of minute volume due to the mixing...

Page 10: ...s See Occlusion troubleshoo ng scenarios to resolve the occlusion alarm If alarm is not resolved remove pa ent from breathing circuit and apply manual breaths with a manual resuscitator Discon nue the...

Page 11: ...ssure limits Assess pa ent status Check the oxygen tubing connec on Check the oxygen source Decrease the Low O2 alarm threshold If alarm is not resolved remove pa ent from breathing circuit and apply...

Page 12: ...e circuit exhala on valve and pressure line for an occlusion Review the ven lator se ngs and alarm limits Press the alarm reset bu on If alarm is not resolved remove pa ent from breathing circuit and...

Page 13: ...for leaks Assess the High Tidal Volume Limit alarm se ng Check the pa ent circuit for leaks or obstruc ons Check the external PEEP valve for malfunc ons Assess the Low Minute Volume alarm se ng Check...

Page 14: ...reset alarm by selec ng the alarm Reset bu on If proceeding with mandatory ven la on select mandatory mode and con rm mandatory ven la on parameters Manually reset the apnea alarm by selec ng the alar...