• Remove pa�ent from breathing circuit and apply
manual breaths with a manual resuscitator.
Discon�nue the ven�lator and send for service.
• Obtain alterna�ve ven�la�on, and remove ven�lator
from use and request service
• Assess pa�ent status
• Review parameter se�ngs for flow rate, inspiratory
�me respiratory rate and �dal volume / pressure limits
• Assess pa�ent status
• Check the oxygen tubing connec�on
• Check the oxygen source
• Decrease the Low %O
2
alarm threshold
• If alarm is not resolved remove pa�ent from breathing
circuit and apply manual breaths with a manual
resuscitator. Discon�nue the ven�lator and send for
service
• If alarm is not resolved remove pa�ent from breathing
circuit and apply manual breaths with a manual
resuscitator. Discon�nue the ven�lator and send for
service
Alarm Troubleshoo�ng
9
When you see
this alarm...
Do This...
Page 11 of 14
AND