SENSIRION Winnipeg Ventilator Quick Reference Manual Download Page 3

Circuit Setup

2

Page 3 of 14

General Informa�on 

To 

change between modes

, push and 

hold

 the desired 

mode bu�on 

for 2 seconds

. This is intended to prevent 

uninten�onal mode selec�on. 

To 

TURN ON

 the ven�lator, push the power bu�on on the 

back of the ven�lator 

To

 TURN OFF 

the ven�lator, place the ven�lator in 

standby mode, then push and 

hold

 the power bu�on 

at the back of the ven�lator 

for 3 seconds

Pa�ent

Inspiratory Port (with internal airway pressure tap)

HME With Filter

Pa�ent Interface

Airway Pressure Tap

Exhala�on Valve
External PEEP Valve

2 X 72 in (183 cm) Hose

Oxygen Source 

(50 psi / 345kPa)

O

2

Expiratory Line Port

Airway Pressure Port

Oxygen Input

Summary of Contents for Winnipeg Ventilator

Page 1: ...Pressure Leak Test Ideal Body Weight MALE 5 Ideal Body Weight FEMALE Full IFU www CEVen lators ca Call Toll Free 1 833 229 1036 Circuit Setup 6 Inspiratory Time 7 I E 9 Alarm Troubleshoo ng Quick Refe...

Page 2: ...ways perform the Pre Use Pressure Test BEFORE connec ng a new circuit to a pa ent In PSV mode the expiratory trigger sensi vity is xed at 10 L min or 25 PEAK inspiratory ow Set the Low Inspiratory Pre...

Page 3: ...on the back of the ven lator To TURN OFF the ven lator place the ven lator in standby mode then push and hold the power bu on at the back of the ven lator for 3 seconds Pa ent Inspiratory Port with in...

Page 4: ...ltaneously for 2 seconds Verify that the Pressure graph reaches 50 cmH2 O See the Insp Tidal Volume display for test status and result If the system fails the pressure test check the system for leaks...

Page 5: ...2 62 157 54 6 218 273 328 382 437 5 3 63 160 56 9 228 285 341 398 455 5 4 64 163 59 2 237 296 355 414 474 5 5 65 165 61 5 246 308 369 431 492 5 6 66 168 63 8 255 319 383 447 510 5 7 67 170 66 1 264 3...

Page 6: ...2 5 2 62 157 50 1 200 251 301 351 401 5 3 63 160 52 4 210 262 314 367 419 5 4 64 163 54 7 219 274 328 383 438 5 5 65 165 57 228 285 342 399 456 5 6 66 168 59 3 237 297 356 415 474 5 7 67 170 61 6 246...

Page 7: ...0 73 0 66 0 60 0 55 0 51 0 47 0 44 0 41 0 39 0 37 0 35 0 33 0 31 0 30 0 29 0 28 600 1 80 1 44 1 20 1 03 0 90 0 80 0 72 0 65 0 60 0 55 0 51 0 48 0 45 0 42 0 40 0 38 0 36 0 34 0 33 0 31 0 30 650 1 95 1...

Page 8: ...Results in impossible se ng where respiratory period is less than inspiratory me 12 14 16 18 20 22 24 26 28 30 35 9 0 7 6 6 5 5 7 5 0 4 5 4 0 3 6 3 3 3 0 2 4 7 3 6 1 5 3 4 6 4 0 3 5 3 2 2 8 2 6 2 3 1...

Page 9: ...50 40 30 20 0 20 40 60 80 100 120 140 160 180 200 5 10 15 20 7 5 25 L min L min L min L min L min L min O 2 se ling me delay on the Winnipeg Ven lator is a func on of minute volume due to the mixing...

Page 10: ...s See Occlusion troubleshoo ng scenarios to resolve the occlusion alarm If alarm is not resolved remove pa ent from breathing circuit and apply manual breaths with a manual resuscitator Discon nue the...

Page 11: ...ssure limits Assess pa ent status Check the oxygen tubing connec on Check the oxygen source Decrease the Low O2 alarm threshold If alarm is not resolved remove pa ent from breathing circuit and apply...

Page 12: ...e circuit exhala on valve and pressure line for an occlusion Review the ven lator se ngs and alarm limits Press the alarm reset bu on If alarm is not resolved remove pa ent from breathing circuit and...

Page 13: ...for leaks Assess the High Tidal Volume Limit alarm se ng Check the pa ent circuit for leaks or obstruc ons Check the external PEEP valve for malfunc ons Assess the Low Minute Volume alarm se ng Check...

Page 14: ...reset alarm by selec ng the alarm Reset bu on If proceeding with mandatory ven la on select mandatory mode and con rm mandatory ven la on parameters Manually reset the apnea alarm by selec ng the alar...

Reviews: