background image

MI00353-2-EN

ENGLISH 4/5

3.8 Dimensions

5.0 ELECTRICAL CONNECTIONS

Remove power supply before wiring.

 

5.1 Output current loop 4 – 20 mA 

A  maximum  voltage  supply  of  28  V 
can be connected directly to the loop 
4 – 20 mA of the T201DC100.

3.7 Standards

Standards

EN61000-6-4 (electromagnetic emission, industry).

EN64000-6-2 (electromagnetic immunity, industry).

EN61010-1 (safety).

Notes:
- Use with copper conductor.
- Use in Pollution Degree 2 Environment
- Power Supply must be Class 2

4.0 PRELIMINARY INSTRUCTION FOR USE

The  T201DC100  can  be  installed  in  any  position  and  place  in  accordance  to  expected 
environmental conditions. Use the included holder bracket when fixing to a DIN rail.

High-strength static magnetic fields may change the output value: keep away from 
permanent magnets, electromagnets or iron bulks that cause such a modification 
of the surrounding magnetic field; try a different orientation if zero error was greater 
than expected.

+

-

U

= 28 V

max

U

I

mA

N.C.
N.C.

+

97 mm

26 mm

68 mm

36.7 mm 

21 mm

Summary of Contents for T201DC100

Page 1: ...ent loop Series of product T Contents 1 0 Disclaimer 2 0 Description and general specifications 2 1 Description 2 2 General specification 3 0 Technical specification 3 1 Input 3 2 Output and supply 3 3 Accuracy 3 4 Environmental condition 3 5 Connector 3 6 Box 3 7 Reference standards 3 8 Dimensions 4 0 Preliminary use instructions 5 0 Electrical connections 5 1 Output current for loop 4 20 mA 5 2 ...

Page 2: ...uring the continuous current component that flowing through it For its electrical endurance ease of use and compact dimensions the T201DC100 fits every kind of current measurement up to 100 Adc 2 2 General specifications Suitable for solar panels batteries battery charger power units and generic DC load No shunt no power consumption from the measure circuit High accuracy rating class 0 2 Loop powe...

Page 3: ... 35 Hz Allowable peak value 30 120 A Full wave rectified 20 80 A Half wave rectified 10 40 A Over voltage category Bare conductor CAT III 300V Insulated conductor CAT III 600V 3 4 Environmental condition Index protection IP20 Operating temperature 20 C 70 C Humidity 10 90 a 40 C non condensing Storage temperature 40 85 C 3 3 Accuracy Maximum errors Input section 0 1 14 mA Output section 0 05 4 µA ...

Page 4: ...tor Use in Pollution Degree 2 Environment Power Supply must be Class 2 4 0 PRELIMINARY INSTRUCTION FOR USE The T201DC100 can be installed in any position and place in accordance to expected environmental conditions Use the included holder bracket when fixing to a DIN rail High strength static magnetic fields may change the output value keep away from permanent magnets electromagnets or iron bulks ...

Page 5: ... Applicable throughout the European Union and other European countries with separate collections programs This symbol found on your producr or on its packaging indicates that this product should not be treated as household waste when you wish to dispose of it Instead it should be handed over to an authorised collection point for the recycling of electrical electronic equipment By ensuring this pro...

Reviews: