background image

MI00353-2-EN

ENGLISH 3/5

3.2 Output and power supply

Type

Passive current loop 4 – 20 mA

Terminals

Removable screw terminal pitch 5 mm for max 2.5mm² cables

Limits

Internal fault / Over-temperature: 3,5 mA
Under-range / Over renge: 3,6 / 21,0 mA
Correct measure: 3,8 / 20,5 mA

Min. - Max. loop voltage

6 V - 28 V

Other protections

Polarity reversal
Loop current limiting on hardware fault
Over-temperature protection

Range

- Unipolar 0 – 10 A, Bipolar -10 – 0 – +10 A
- Unipolar 0 – 25 A, Bipolar -25 – 0 – +25 A
- Unipolar 0 – 50 A, Bipolar -10 – 0 – +50 A
- Unipolar 0 – 100 A, Bipolare -25 – 0 – +100 A 

AC 

superimposed

 (f 35 

Hz)

- Allowable peak value: -30 – +120 A
- Full-wave rectified: -20 – 80 A
- Half-wave rectified: -10 – 40 A

Over-voltage category

Bare conductor

CAT. III 300V

Insulated conductor

CAT. III 600V

3.4 Environmental condition

Index protection

IP20

Operating temperature

-20 °C – +70 °C

Humidity

10 – 90% a 40 °C non-condensing

Storage temperature

- 40 – +85°C

3.3 Accuracy

Maximum errors

- Input section: 0,1 % + 14 mA
- Output section: 0.05 % + 4 µA

TempCo

< 150 ppm/K

Error due to EMI

< 50 µA, test on bare wire Ø 10 mm

Response time

- Without damping filter: 100 ms
- With damping filter: 600 ms

3.5 Connections

Connections

Removable screw terminal pitch 5 mm

 for loop 4 – 20 mA

Tightening torque 7.0 Lb•inch = 0.8 N•m

21 mm through hole for wire

3.6 Box

Dimension

97 x 68 x 26 mm without screw terminals

Box

PA6, Black color

Summary of Contents for T201DC100

Page 1: ...ent loop Series of product T Contents 1 0 Disclaimer 2 0 Description and general specifications 2 1 Description 2 2 General specification 3 0 Technical specification 3 1 Input 3 2 Output and supply 3 3 Accuracy 3 4 Environmental condition 3 5 Connector 3 6 Box 3 7 Reference standards 3 8 Dimensions 4 0 Preliminary use instructions 5 0 Electrical connections 5 1 Output current for loop 4 20 mA 5 2 ...

Page 2: ...uring the continuous current component that flowing through it For its electrical endurance ease of use and compact dimensions the T201DC100 fits every kind of current measurement up to 100 Adc 2 2 General specifications Suitable for solar panels batteries battery charger power units and generic DC load No shunt no power consumption from the measure circuit High accuracy rating class 0 2 Loop powe...

Page 3: ... 35 Hz Allowable peak value 30 120 A Full wave rectified 20 80 A Half wave rectified 10 40 A Over voltage category Bare conductor CAT III 300V Insulated conductor CAT III 600V 3 4 Environmental condition Index protection IP20 Operating temperature 20 C 70 C Humidity 10 90 a 40 C non condensing Storage temperature 40 85 C 3 3 Accuracy Maximum errors Input section 0 1 14 mA Output section 0 05 4 µA ...

Page 4: ...tor Use in Pollution Degree 2 Environment Power Supply must be Class 2 4 0 PRELIMINARY INSTRUCTION FOR USE The T201DC100 can be installed in any position and place in accordance to expected environmental conditions Use the included holder bracket when fixing to a DIN rail High strength static magnetic fields may change the output value keep away from permanent magnets electromagnets or iron bulks ...

Page 5: ... Applicable throughout the European Union and other European countries with separate collections programs This symbol found on your producr or on its packaging indicates that this product should not be treated as household waste when you wish to dispose of it Instead it should be handed over to an authorised collection point for the recycling of electrical electronic equipment By ensuring this pro...