10-90% response :
Transmission :
Max. transmission error
:
(1)
Resolution :
Thermal drift :
SQRT error
:
(2) (3)
Linearisation error Cylindrical
tank
:
(2)
50 Hz : max 41 ms without filter and 88 ms with filter;
60 Hz : max 35 ms without filter and 74 ms with filter.
Digital Optical
0.08% of the f.s. value for mA or 5 V output
0.07% of the f.s. value for 10 V output
1 mV for voltage output, 2 A for current output
m
Lower than 120 ppm/K
in the range 1..100%: floating point 32 bit
0,05%
MI00102 -E
4
ENGLISH - 2/8
Insulation Voltage :
Protection Index :
Operating Conditions :
Storage Temperature :
LED Signalling :
Connections :
Conductor Section :
Wire stripping :
Standards :
Dimensions, Weight :
1,5 KV between each group of ports
IP20
Temperature -20..+65 °C
Humidity 10..90 % at 40°C (non-condensing)
Altitudine 2000 slm
-40..+85 °C
Input or output out-of-range limiter device triggered or
input saturation Internal fault. Wrong DIP-switches
.
settings.
Spring terminals
0,2..2,5 mm
2
EN61000-6-4/2002 (electromagnetic emission, industrial surroundings)
EN61000-6-2/2005 (electromagnetic immunity, industrial surroundings)
EN61010-1/2001 (safety)
All the circuits must be provided with double insulation from the circuits
under dangerous voltage. The power supply transformer must be built
to compliance with EN60742: “Insulation transformers and Safety
transformers”.
Notes:
- Use with copper conductor.
- Use in Pollution Degree 2 Environment .
- Power Supply must be Class 2.
- When supplied by an Isolated Limited Voltage/Limited
Current power supply a fuse rated max 2.5 A shall be
installed in the field.
6,2 x 93,1 x 102,5 mm, 46 g.
Box :
PBT (black colour)
(1)
No linearisation function enabled.
(2)
Linearisation functions operate only in the 0..100% rated range, whereas for the under-
range and the over-range, the input signal is transferred without any alteration (G=1).
Continuity and monotonic quality of transfer are guaranteed throughout the entire range of
measurement.
(3)
In the 0..1% range, the function is linear with gain G=10 in order to avoid over-
amplification of the noise.