0000137188 Rev. 1.0
12
Press
Esc
to exit any menu and confirm the settings by pressing
Enter.
DISPLAY VIEW WITH THE SYSTEM ON STAND-BY
*If Advanced
Flow=On and during H
2
O
2
dosing, the recirculation pump stops and
blocks the system dosing, when the pump restarts, the H
2
O
2
will restart the cc dosing
calculated from the start.
SHORTCUTS
To access the shortcut menu press the
SET
key for at least 3 seconds when the system is in stand-by.
Display
Settings
Progrma Menu
Exit Save
Press the + and
– keys to choose Save or
NoSave, i.e. to save the settings or not and
press
Enter
to confirm
.
Stand-by display
Operation
Alr
Tm 25°C 8.3pH
Alr 1200cc
Tm 25°C 8.3pH
Alr H
2
O
2
37m
Tm 25°C 8.3pH
The display is divided into 4 parts:
On the upper left hand side the alarm is
shown, when present.
On the upper right hand side three views are
displayed:
o
The first is the stand-by system
o
The second shows the cc countdown
dosed by the pump
o
The third is the waiting time until the
next dosing with the message H
2
O
2
.
On the lower left hand side the temperature
read by PT100 or manually set is shown.
On the lower right hand side the value read
by the pH probe is shown
Display Set
Operation
pH = 7.2
H
2
O
2
= 10cc/m3
The pH- value flashes. It can be modified using the +
and
– keys and confirmed using
Enter.
For the H
2
O
2
value, repeat the same procedure, press
Enter
to
confirm and exit.
1