To maximize your chances of an accurate comparison between meter
and laboratory results, follow a few basic guidelines:
Before going to the lab
·
Perform a control solution test to make sure the meter is working
properly.
·
Do Not eat for at least eight hours before you test your blood.
Take your meter with you to the lab.
·
Performance Characteristics
y-intercept 24.087
correlation coefficient
(
R
²)
24.087
Precision:
Average
concentrarion(umol/L) 272.6 430.6 532.6 709.4 928.9
SD(umol/L)
16.8 20.3 25.4 28.8 37.7
CV(%)
6.2% 4.7% 4.8% 4.1% 4.1%
Slope
0.9493
42