Testing
your
blood
glucose
Testing
①
With meter off or in setting mode and memory mode,
insert a test strip to enter into testing mode.
Note: If you do not start the test within three minutes, the
meter will turn off. To restart your meter, take out the
unused test strip and reinsert it into the meter.
Important: The meter should only be used with BS-102
test strips. Using other test strips with this meter can
produce inaccurate results.
②
System check screen
Note:
•
If appears, it indicates the operating temperature is
over range. Place the system in operating temperature
room for half hour. Then test again.
•
If appears, it indicates the battery is almost out
of power. Replace the battery now.
③
The blood drop symbol flashes on the display screen.
Your meter is now ready to apply blood sample.
20
AVG
OK
mg/dL mmol/L
MEM
M
D
AM
PM
The backlight will be lit every time the meter is
turnedon. The screen will briefly display all content
to confirm that the display is working properly.