9
2.3 Electrical Specifications
Accuracy is given as ±(% of r number of digits) from 18 °C to 28 °C, with relative humidity up to 70%.
Temperature coefficient: 0.1 x specified accuracy/°C (t <18 °C or t >28 °C).
DC Current - Manual ranging
Range
Resolution
Accuracy
Overload Protection
400A
100mA
( 1.5% + 5 dgt)
1000A
1A
0-600A
( 1.5% + 5 dgt)
600-800A
( 2.5% + 5 dgt)
800-1000A
( 3.5% + 5 dgt)
1200A for 60 s maximum
Specifications given from 5% to 100% of the range. Accuracy given after having performed auto-zero of the clamp.
AC Current - Manual ranging
Range
Resolution
Accuracy 50-60Hz
Accuracy 60-400Hz
Overload
Protection
400A
100mA
( 1.5% + 5 dgt)
( 3.0% + 5 dgt)
1000A
1A
0-600A
( 1.5% + 5 dgt)
600-1000A
( 2.0% + 5 dgt)
0-600A
( 3.0% + 5 dgt)
600-1000A
( 3.5% + 5 dgt)
1200A for 60 s
maximum
Specifications given from 5% to 100% of the range.
Crest factor :
3 max. Add 1% to accuracy for crest factor between 2 and 3
.
Conversion :
TRUE RMS.
Summary of Contents for MW3105
Page 1: ...MW3105 DIGITAL CLAMP MULTIMETER...
Page 2: ...2 M MW3105 A 01...
Page 6: ...6 Figure 1...
Page 27: ...27...