7
3.2
Connection
Connect the cables from OCC (P17 Figure 3-3 #1, SHB CONTROL and P18 Figure 3-3 #2, SHB POWER) to the
mating connectors on the rear of the SHB. The SHB CONTROL cable connects to the 28 pin connector on the
SHB (left side of rear panel, looking from rear of unit). The SHB POWER cable connects to the 8 pin connector on
the SHB (right side of rear panel, looking from the rear of unit). If the optional foot-switch has been ordered,
connect it to P19 (Figure 3-3 #3) on the OCC.
Verify that the voltage required is appropriate for your power line voltage. The required voltage can be determined
by examining the power input connector assembly (P5 Figure 3-3 #4) on the rear of the OCC. Count-A-Paks are
designed for 115V or 230V operation, although they will operate over a range of 100-130V or 200-260V. There
are 4 small holes on the right-hand end of this connector assembly, which are labeled 100V, 120V, 220V, and
240V. There will be a small white indicator pin in one of these holes (see Figure 7-1). The voltage indicated
beside that marker is the line voltage for which the OCC is configured. The Count-A-Paks are typically shipped
with this set to 120V or 240V position. The 100V and 220V positions are used only for very low line voltages.
Additionally, all units which require 200V to 260V will be labeled "220 VOLTS" on both the OCC and the line cord
(for further information see Section 7.3).
Figure 3-3 Connections