![Seedburo 801-10/B Operating And Service Manual Download Page 20](http://html.mh-extra.com/html/seedburo/801-10-b/801-10-b_operating-and-service-manual_1232856020.webp)
19
"
SENS. OK?
"
Verify current Sensitivity setting with ENTER/YES, or change setting utilizing one of several
methods. The setting can be changed using the
up/down arrows. It can also be changed entering a new value via the numeric keypad
followed by ENTER/YES. Also, if CLEAR/NO is pressed, the message "MODIFY NOW!!"
will appear and a new value can be entered via the numeric keypad followed by
ENTER/YES. Note: The Sensitivity can be adjusted during counting via the up/down
arrows.
"
BOWL EMPTY
"
This message appears when a seed has not been counted in several minutes. This is the
normal termination in TOTAL and BATCH REPEAT modes.
"
NO BAG
"
This indicates that a bag must be attached to one of the chutes, using the bag retainer clip,
prior to starting, or a bag was removed during a cycle. This function may be disabled via
the "CHK SETTNGS" function in the main menu.
"
FINISHED
"
This message appears when the counting is complete in SINGLE BATCH or MULTIPLE
BATCH modes.
"
COUNT OK?
"
Verify current Count setting with ENTER/YES, or change setting utilizing one of several
methods. It can be changed by entering a new value via the numeric keypad followed by
ENTER/YES. Also, if CLEAR/NO is pressed, the message "MODIFY NOW!!" will appear
and a new value can be entered via the numeric keypad followed by ENTER/YES.
"
BATCHES OK?
"
Verify the current number of batches in this Group with ENTER/YES, or change setting
utilizing one of several methods. It can be changed by entering a new value via the numeric
keypad followed by ENTER/YES. Also, if CLEAR/NO is pressed, the message "MODIFY
NOW!!" will appear and a new value can be entered via the numeric keypad followed by
ENTER/YES.
"
GROUP OK?
"
Verify the current settings of Terminal Count and Batches in this Group with ENTER/YES,
or change settings by pressing CLEAR/NO.
"
MODIFY NOW!!
"
Enter a new value for the current parameter via the numeric keypad followed by
ENTER/YES.
"
PRESS START
"
Press START to initiate counting.
"
PRESS STOP
"
Press STOP to interrupt the counting.
"
CHK SETTNGS?
"
Press ENTER/YES to modify the current default settings of bag detector operation, line
frequency, gate delay, and seed tone operation. Press CLEAR/NO to return to the
"SELECT MODE" function.
"
REVISION X.X
"
This is displayed immediately after power-up prior to calibration. The number which
appears in place of X.X is the software revision number which should be provided to obtain
support.
"
BAG DET. ON?
"
Pressing ENTER/YES enables the Bag Detect circuit. CLEAR/NO inhibits its operation.