High Frequency Generators - Veterinary Systems APR-VET
Operation
OM-0118R0
13
3.1
RADIOGRAPHY AND GENERAL CONTROLS
The Radiographic module consists of: Power On / Off, Bucky / No Bucky
selection, displays and controls of radiographic parameters, Focal Spot, Status
and Error indicators.
3.1.1
POWER ON / OFF
ON:
The Generator is turned ON by pressing this push-button. This starts a
power-up routine and the Console shows the software version (a.e. P05.3 =
Vers.5, Rev.3). After the power-up routine the last workstation used will be
automatically selected.
OFF
: The System is turned OFF by pressing this push-button.
IN THE EVENT OF AN EMERGENCY, ISOLATE THE X-RAY
SYSTEM FROM THE MAINS POWER BY FORCIBLY
DEPRESSING THE X-RAY ROOM “EMERGENCY OFF
SWITCH” (USUALLY A RED MUSHROOM-SHAPED SWITCH).
THIS SWITCH SHOULD BE LOCATED ON OR NEAR TO THE
X-RAY ROOM ELECTRICAL CABINET, USUALLY PLACED
NEAR TO THE GENERATOR CONTROL CONSOLE. MORE
THAN ONE OF THESE SWITCHES MAY BE PLACED AROUND
THE ROOM FOR GREATER ACCESSIBILITY.
3.1.2
BUCKY / NO BUCKY SELECTION
BUCKY / NO BUCKY
: Press this push-button to select the Bucky configured in
the system (push-button lighted), or to select “
No Bucky
” (push-button off).