High Frequency Generators - Veterinary Systems APR-VET
Operation
OM-0118R0
23
Each Thickness Range of a Body Regions has its own slope of the
kVp line. When it is required to move all the line, change and store
both APR techniques at maximum and minimum thickness in the
selected range.
1
5
10
60
70
80
kVp
cm
1
5
10
60
70
80
kVp
cm
84
kVp line of the pre-programmed APR technique
1
5
10
60
70
80
kVp
cm
84
First modification of the kVp line
Final kVp line of the APR technique
64
G
If the thickness value (cm) of the stored APR technique is not the
minimum or maximum value of its Thickness Range, information
about the selected kVp is not stored.
G
Independently of the thickness value (cm), the new mAs value (or
mA and Exposure Time) is stored in all cases for the selected
Thickness Range. Also, information about AEC selections is
stored in all cases (when AEC is installed) for the selected
Thickness Range.