3.4
Getting Started (Insert / Replace Accumulator)
The accumulator compartment is located in the back of the
Easyflex Com
.
Make sure that all test leads have been disconnected before opening the
accumulator compartment. Switch the device off in order to retain all data.
For inserting or replacing the accumulator open the lid and pull out the
accumulator, if there is one. Insert the charged NiMH replacement
accumulator with its contacts facing the receptacle of the compartment.
Make sure that the accumulator is inserted properly, otherwise you would
be unable to close the lid. Now the
Easyflex Com
is ready for operation.
3.5
Switching On and Off
Press key
to switch the
Easyflex Com
on. During run-up the
manufacturer’s logo will be shown for about 2 seconds.
After run-up has been completed, the
Easyflex Com
can be switched off any
time by pressing key
. When being switched off the device saves all
device settings, so the same operating status is restored when the
instrument is switched on next time.
The
Easyflex Com
cuts back its function automatically some time after a key
has been pressed last time in order to save accumulator power:
•
The device is switched off after 360 sec,
•
The back light of the display is switched off after 60 sec,
•
The reduction of measuring rate is disabled
These default values as mentioned above can be changed in menu Power
Supply.
Device on/off
off
10 sec ... 20 min
in steps
of 10 sec
Light on/off
off
5 sec ... 10 min
in
steps of 5 sec
Meas.
rate
reduction
off
1 sec ... 1 min
in
steps of 1 sec
Key operations: 2nd; Menu; rotary encoder to Supply menu; Mark; rotary
encoder for Vertical selection; Mark; use rotary encoder to set Time; Mark;
Display