
Putting in contact 2 metals from different material generates unwanted voltages
(EMF) proportional to the ambient temperature.
Two methods exist to cancel this problem: the Zero method and the Average method.
The ZERO method:
Φ
Select the ZERO line
–
ON
= Zero method operating
–
OFF
= Zero method not operating
Φ
Enter the selection with the Enter key.
The AVERAGE method:
Φ
Select the AVERAGE line
Φ
Press on the RIGHT arrow or on the enter key.
measurements to get the resistance calculation. Select with the UP and DOWN
arrows:
–
ON
= Average method operating
–
OFF
= Average method not operating
Φ
Enter the selection with the Enter key.
RANGE OFF
CURRENT
ON
ZERO
AVERAGE
>
LOG
LIMITS
ESC
RANGE OFF
CURRENT
ON
ZERO
>
AVERAGE
LOG
LIMITS
ESC
When measuring low resistances, consequently low voltages, the accessories and
4-1-3 Thermal EMF Error:
Both methods are available with the DO7010.
Φ
The Zero method consists of a voltage measurement without current, then a
Φ
The Average method consists of a voltage measurement with a positive
resistance calculation. Select with the UP and DOWN arrows:
Φ
Press on the RIGHT arrow or on the enter key.
voltage measurement with current, and subtraction of EMF voltages before
current, then a voltage measurement with negative current, and average of the 2
specimen under test conductive metals can introduce significant errors.
Operating Instructions
Digital Microhmmeter Type DO7010
28
•
•
•
•
•
•
•
•
Summary of Contents for CROPICO DO7010
Page 44: ...610021 Rev 3...