4-1-1 Measurement range selection:
Φ
Select the RANGE line
Φ
Press on the RIGHT arrow or on the enter key .
6000m
Ω
: from 0001 to 6000 m
Ω
600m
Ω
: from 000.1 to 600.0 m
Ω
60m
Ω
: from 00.01 to 60.00 m
Ω
6m
Ω
: from 0.001 to 6.000 m
Ω
Φ
Enter the selection with the Enter key.
4-1-2 Measurement current selection:
Φ
Select the CURRENT line
Φ
Press on the RIGHT arrow or on the enter key.
Φ
Select the measurement current with the UP and DOWN arrows:
HI = 10A for the 6 to 600m
Ω
ranges and 1A for the 6000 m
Ω
range.
LO = 1A for the 6 to 600m
Ω
and 0.1A for the 6000 m
Ω
range.
Φ
Enter the selection with the Enter key.
RANGE LO
CURRENT
>
HI
ZERO
AVERAGE
LOG
LIMITS
ESC
RANGE
>
6000m
Ω
CURRENT 600m
Ω
ZERO
60m
Ω
AVERAGE 6m
Ω
LOG AUTO1
LIMITS
AUTO2
ESC
Φ
Select the measurement range with the UP and DOWN arrows:
AUTO2 : automatic searching range from the last measurement range.
AUTO1 : automatic searching range from the highest range (6000m ).
Ω
Operating Instructions
Digital Microhmmeter Type DO7010
27
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•