Earth Bond Test
Chapter 7 Specifications
Earth Bond Test
Test Voltage *
.....................................................6V nominal (no load)
Test Current *
..........…...……...4A, 10A, 25A selectable (into s/c
load)
Range
.......................................................................... 40m
Ω
- 19.99
Ω
Resolution
.............................................................................…..0.01
Ω
Accuracy
................................................. +/- 5% of reading, +/- 2 digits
Pass
Levels
...........................................................................selectable
Misc
...........................................….4-wire measurement, floating earth
Earth Screen Test
Test Voltage *
............................................. 100mV nominal (no load)
Test Current *
............................................……..100mA (into s/c load)
Range
.......…………………………………………………40m
Ω
-
19.99
Ω
Resolution
............................................................................…...0.01
Ω
Accuracy
………………….+/- 5% of reading, +/- 2 digits up to 5.00
Ω
Pass
Levels
...........................................................................selectable
Misc
...........................................….4-wire measurement, floating earth