How to print Barcode Labels and Testcodes
Default is 0%.
POWER POLARITY
This option will allow the tester to perform the power tests in one of
three ways.
INPUT POLARITY - perform the power tests with the same mains
polarity as the Supernova Plus is powered.
AUTO CORRECTED - perform the power tests with the mains correct
polarity even if the Supernova Plus is powered from a socket which
has the mains polarity reversed.
INPUT & REVERSED - this option will perform two power tests for
every one selected, one with the correct mains polarity and one with
the reverse mains polarity.
Default is INPUT POLARITY.
ON TEST FAILURE
This gives the User the option of having a menu for a failure within an
automatic sequence. This menu will give you the option of retrying a
failed test or appliance.
Default is FAIL MENU.
COMMENTS
This option allows the User to determine if the Comments field will be
available at the end of a test.
ALWAYS – Comments always available at the end of test
ON PASS – Comments only available if the Appliance passed the
test.
ON FAILURE - Comments only available if the Appliance failed the
test.
NEVER – Comments will not be available at the end of the test.
Default is ALWAYS
APLNCE & CMNTS
This allows the user to set the number of characters available for the
appliance number and comments.
Options are:
APP = 10,
CMNTS = 20
APP = 15,
CMNTS = 20
APP = 10,
CMMTS = 10
Default is APP = 10,
CMNTS = 20