XFp Extracellular Flux Analyzer User Manual 

_____________________________________________________________________________________________________________________ 

 

Suitable Locations for the XFp Analyzer 

The XFp Analyzer is designed for laboratory use.  The internal environment for the XFp Cell Culture 
Miniplate is controlled to a preset temperature; therefore, laboratory room temperature must be 
maintained within the range listed in the specification table in Section 2.  The miniplate tray 
temperature can be monitored using the status display on the upper right portion of the user interface. 

The XFp Analyzer uses optical detection technology to measure extremely low levels of fluorescent 
emission from analyte sensors.  Therefore, although the instrument has been designed to shield room 
light, excessive light (such as direct sunlight) should be avoided. 

 

 

Safety Considerations

 

WARNING: 

The protection provided by this instrument may be compromised if it is used in a manner not 

specified by Seahorse Bioscience.

 

 
Safe operation of the XFp Analyzer requires that all covers be securely attached and plate tray door is 
closed. This also prevents heat loss and system cooling, which can affect data quality.  
 
The door opens automatically when the tray is extended, allowing the operator to insert or remove the 
assay consumables.  The operator must exercise caution during the loading of the miniplate/cartridge in 
order to avoid the possibility of a pinch hazard.  After the miniplate/cartridge is placed securely on the 
tray, the operator’s hand must be removed from the area of the tray before continuing the assay.  After 
the command is given to continue the assay via the user interface, the tray will move slowly back into 
the instrument and the door will close.  An optical sensor is used to determine the status of the door. 
 
The  XFp  Analyzer has heaters  around  the miniplate  that  maintain  a stable temperature. Typically, the 
temperature  will  be  maintained  at  37°C,  as  monitored  by  temperature  sensors  and  controllers 
embedded in above the tray. A thermal fuse will disable the heater should it reach an abnormally high 
temperature. 
 
The user shall not replace the power cord provided with any other power cord that is rated at less than 
what is specified in Section 2, dependent on the power mains of the country the instrument will be used 
in. 

 

 

 

 

Summary of Contents for XFp

Page 1: ...XFp Extracellular Flux Analyzer User Manual Part 102745 400 Rev A...

Page 2: ...d N Billerica MA 01862 Submit technical support questions or comments about our documentation to support seahorsebio com Phone support Global United States United States Toll Free United Kingdom Germa...

Page 3: ...ty Considerations 7 4 Basic Operation of the XFp Analyzer Power and Warm Up 8 Welcome Screen on First Time Power Up 8 Communications and Network Connection Setup 8 Turning the XFp Analyzer Off 8 5 Nav...

Page 4: ...concentration drops approximately 20 30 and media pH declines approximately 0 1 0 2 pH units The measurement is performed using optical fluorescent biosensors embedded in a disposable cartridge that i...

Page 5: ...43 cm x 58 cm Weight 33 lbs 15 Kg Power Requirements 100 240V AC 50 60Hz 6A 3A Power cord rating 3 wire grounded AC power cord rated 10A or greater Power fuse ratings 250V 5A Time Lag 2 fuses 5mm X 20...

Page 6: ...on Unpacking and Component Identification Upon receipt immediately check each box for damage Report any shipping damage to the transportation company and Seahorse Bioscience using the contact informat...

Page 7: ...vers be securely attached and plate tray door is closed This also prevents heat loss and system cooling which can affect data quality The door opens automatically when the tray is extended allowing th...

Page 8: ...Navigating the XFp Analyzer Software XFp Analyzer Communications and Network Connection Setup The transfer of template and result files between the XFp Analyzer and an analysis computer running Wave c...

Page 9: ...pear and allow selection and running of a template from the networked location ii If a USB thumb drive containing a valid XFp template has been inserted into the instrument a USB tab will appear Templ...

Page 10: ...uide for XFp Analyzer for more information about measurement cycles 4 Review Summary and Start Assay a Review the General Information and Errors and Warnings sections to verify that the settings are a...

Page 11: ...Calibration of the sensor cartridge which takes approximately 20 minutes Note Although the sensor calibration may appear to be complete after 15 minutes the instrument will be busy for a few more minu...

Page 12: ...uching the Options button different data can be displayed 1 Show StdDev turns on display of error bars for the each Group 2 Level Data shows the concentration data from which the rate data is derived...

Page 13: ...te 7 After the assay is complete the data will automatically be saved to a USB flash drive if inserted the local drive or the networked drive if configured A message will appear indicating the locatio...

Page 14: ...which the instrument is located 2 Backup Assay Results a Backup Assay Results saves a copy of each assay performed within the previous 60 days on the local storage within the XFp Analyzer i Insert a U...

Page 15: ...work access The Ethernet cable port is located on the lower backside of the XFp Analyzer b Wireless Network A wireless USB adapter can be plugged into one of the available USB ports recommended on low...

Page 16: ...Overall System Communication Test Verifies all system modules are communicating appropriately Voltage Test Ensures voltage levels are within specified ranges for operation Motor Test Verifies the move...

Page 17: ...XFp Analyzer is designed to require minimal cleaning and maintenance All consumables are disposable and none of the instrument components are exposed to the cell plate at any time preventing cross con...

Page 18: ...ion is contained on the Help screen 1 Support Worldwide Seahorse Support contact information 2 Version of Software Hardware Currently installed software and firmware versions on XFp Analyzer If the XF...

Page 19: ...nd installed on any personal computer via the Seahorse Bioscience website http www seahorsebio com support software update php Refer to the Wave User Guide for XFp Seahorse Document 103244 400 for a d...

Page 20: ...internal Wi Fi adapter in the instrument The Ethernet RJ 45 jack is located at base of the instrument in the back The XFp Analyzer can be connected to any Microsoft Windows compatible network and the...

Page 21: ...al IP Address Assignment 1 Remove the green check from the Obtain IP Address Automatically check box 2 Manually enter the information provided by the local IT department for the instrument 3 After set...

Page 22: ...To Setup 10 Click the Ping button in the Wired Network Connection Settings screen 11 If there is a Pass message and icon the wireless setup is complete a If it fails confirm the information in the wir...

Page 23: ...t can be automatically set Time Zone Setup 1 From the Home screen click Settings and then click Go To Setup 2 Press the right arrow at the bottom right four times The Time Zone tab will appear 3 Selec...

Page 24: ...ail result delete the contents of every field on this page and click Apply The templates page should now be displayed quickly Instrument gives a load error message and it is not clear whether a cartri...

Page 25: ...ates Toll Free 800 671 0633 option 3 Global United States 1 978 671 1600 option 3 United Kingdom 0800 096 7632 Germany 0800 180 6678 Europe 45 31 36 98 78 Consumables Order To order XFp FluxPaks inclu...

Reviews: