SciQuip Oven-140HT User Manual Download Page 8

 
 

 

Change/Confirm button:   

Press this button to display clock time. Hold on this button for 3 seconds to change the 
parameter, and press this button again to confirm the change. 

Increase button:  

Press this button, the parameter will increase by one digit, the parameter will keep 
increasing while this button is held on. 

 Decrease button:  

Press this button, the parameter will decrease by one digit, the parameter will keep 
decreasing while this button is held on . 

 Program button:  

Press this button to set the program control parameter. 
 

Temperature button:  

Press this button to display the preset temperature, press this button again to confirm 
the change. 

 Time button:  

Press this button to display the operating time, press this button again to display the 
preset time. 

Operating button: 

Under constant temperature control mode, press this button to start operation and press 
it again to stop  operation. Under programme control mode, press this button to start 
operation and press it again to keep the status, hold on this button for 3 seconds to stop 
operation. 
 

Temperature exceeding alarm indicator:  

When real temperature exceeds the upper or lower temperature limit, alarm will be 
activated, this indicator  will be light and the buzzer will keep alarming. 
 

Heating indicator:  

  When the heater is connected with power supply, this indicator will be light. This 

indicator will twinkle when the real temperature close to the preset value. 

 Real temperature Indicator:  

When the screen is displaying the real temperature, this indicator will be light. 

 Temperature setting Indicator:  

 When the screen displays the preset temperature, this indicator will be light. 

 

 Time indicator:  
 

When screen displays the operating time, this indicator will be light, twinkles when the 
screen displays preset time. 

 

 Operating status indicator:  
 

When the equipment is well functioned, this indicator will be light. 
 
 
 

HEATING 

SV 

TM 

RUN 

Summary of Contents for Oven-140HT

Page 1: ...User s Manual This Manual applies to Oven HT series horizontal airflow from back Oven 80HT 140HT 230HT Oven HT series Blast Air Oven...

Page 2: ...ration of specimens and the drying sterilization baking operations of specimens like food electronic components building material etc SciQuip thank you for your trust in its product Since the date of...

Page 3: ...mable or explosible object inside the chamber 8 When in operation the window of the equipment has a relative high temperature so any physical contact with it should be avoided 9 To avoid burn do not f...

Page 4: ...door heavily will probably leads to the damage of the equipment 8 When in operation do not open the door indiscriminately which will affect the temperature inside 9 The exterior of equipment must be...

Page 5: ...4 5 Setting of the Control Mode 4 6 Setting of the parameter under the constant temperature control mode 5 7 Setting of the parameter under the programme control mode 7 8 Setting and Application of ot...

Page 6: ...as power cut alarm sensor failure alarm and upper and lower temperature limit exceeding alarm 7 Highly responsive thermal sensor and environmental scanning microprocessor chip 8 Three dimensional heat...

Page 7: ...ity 2 5 max temperature Display LED Alarm Enabled Timer min 1 999 Settings Digital Grids Included 2 max13 2 max17 2 max21 Grid Size mm WxD 319 330 385 405 475 485 Chamber Dimensions mm WxDxH 375 365 5...

Page 8: ...art operation and press it again to stop operation Under programme control mode press this button to start operation and press it again to keep the status hold on this button for 3 seconds to stop ope...

Page 9: ...rature Then the equipment is in the state of readiness 5 Setting of the control mode This equipment has two operating mode i e constant temperature control and programme control users can choose one a...

Page 10: ...e red TM indicator turns on Press the Time button again to display the preset timing parameter minute one decimal the red TM indicator twinkles Press the Increase or Decrease button can change the par...

Page 11: ...the setting of constant temperature control mode Below is a graph showing the relation under normal condition between time and temperature when temperature keeps increasing It can be referred to when...

Page 12: ...ramme segment 3 P SL1 The ending temperature of the slope part of the first programme segment 4 P td1 Time of temperature retaining of the temperature retaining part of the first programme segment 5 P...

Page 13: ...m enters into the temperature retaining segment the setting temperature is the ending temperature of the slope segment The ending temperature of the last slope segment is the starting temperature of n...

Page 14: ...rogram begin operation from the segment displayed on the screen and take the current real temperature as the starting temperature 6 The setting and application of the parameter Hb feedback value Param...

Page 15: ...educe to 0 indicator HOLD turns off operation begins the red operating indicator RUN turns on Under programme control mode operation of programme segment begins Simplified instruction on the setting o...

Page 16: ...button for 3 seconds screen displays CIP a second later 0 press the Increase button to input the password 4 press the Change Confirm button again screen displays the hour value press the Increase or D...

Page 17: ...ting time reaches the preset value heating output will be stopped the operating indicator turns off At this time the power control transformer is still connected to power supply so turn off the main s...

Page 18: ...3 Malfunction of heating system occurs Wait a minute and observe for a while Refer to the operating procedure and change or re set parameter Notify the manufacturer for repair service Real temperature...