4 Maintenance
Page 19
STAT
MATIC
smart 1,2,3
EN
Data backup of the log file
Switch off the device
Insert the USB stick on the back of the device
Turn on the device
Press the
button. "Export log data" will appear on the display.
Use the
button to start exporting the log files to the USB stick.
When the process is finished, the display shows the following: Log data
export complete.
Use the
button to return to the menu.
Switch off the device
Remove the USB stick.
•
If the USB port is to be extended using a USB cable, this cable must be equipped with a ferrite core. The cable must be
inserted with the ferrite core facing the device side.
•
Please note that the data on the USB stick must be copied to a suitable permanent storage device. Please check that
the data on the storage medium can also be read correctly.
Deletion of the log file on the STATMATIC smart
Deletion of the log file causes the data to be permanently lost. Please create a backup of the log file first and check
whether the data on the USB stick can also be read correctly.
Turn on the device
Press the
button. "Export log data" will appear on the display.
Press the
button 6x until "Delete log file" is highlighted in black on the display.
"Deletion" is selected using the
button.
Press the
button once more. The deletion of the log data will be performed.
Use the
button to return to the menu.
Press the
button to return back to the "Cl Maintenance" main menu
Emptying the media system
This is only required if the system is to be emptied completely - for example for transport purposes or longer standstill
periods (>30 days).
First, empty the containers completely using drain hoses. Then, tilt the device to the rear by approx. 45 degrees in order
to also empty the residual liquids located at the bottom of the container using the drain hoses. Media can only be re-used
again if they have been kept in their original bottles (mind the best before dates on the bottom or on the sides of the
original bottles).
Turn on the device
Press the
button. "Export log data" will appear on the display.
Press the
button 2x until "Empty media system" is highlighted in black on the display.
Using the
button, the "Empty media system" menu is selected.
The media system will be automatically emptied. During this time, a clock will run backwards until 00:00.
Press the
button to return to the "Cl Maintenance" main menu
USB connection
(rear panel of the device)