3 Operation
Page 13
STAT
MATIC
smart 1,2,3
EN
The processes "Clean only" and "Maintenance only" can also be carried out individually.
"Maintenance only" program - for regular and contra-angle handpieces and turbines
Turn on the mains switch.
LED
(Stand-by) lights up green. "Cl
Maintenance" will appear on the display.
Open the front door.
Place regular and contra-angle handpieces and turbines
on the adapters provided for this purpose (instruments
must snap into place). A frame will appear on the display
around the station number per occupied slot.
Close the front door.
Press the
button 1x. "Maintenance only" will appear on
the display.
Push the
(Program Start) button on the control
membrane keyboard.
Program runs automatically. During the maintenance
program, a clock will count backwards until 00:00 on the
display.
All attached regular and contra-angle handpieces and
turbines are maintained for inside in turn. After a
successful process, the program slots will be displayed
inversely. The display shows: "Maintenance successful".
Open the front door.
Remove regular and contra-angle handpieces and turbines
by pressing the locking button.
Pull off turbines downwards (press support lever).
The frame around the maintenance slot number will
disappear from the display.
•
Wipe off the instruments using a damp cloth soaked with disinfectant. For Germany we recommend using disinfectants
listed by the VAH (such as SciCan OPT
IM
BLUE).
•
Immediately following this treatment, the instrument must be treated further in accordance with the manufacturer's
instructions on reprocessing (such as disinfection/sterilization in the STAT
IM
).
•
Always perform a function test before using the instruments on patients. Some liquid may come out.
"Clean only" program - for regular and contra-angle handpieces and turbines
Turn on the mains switch.
LED
(Stand-by) lights up green. "Cl Maintenance" will appear on the display.
Open the front door.
Place regular and contra-angle handpieces and turbines on the adapters provided for this purpose
(instruments must snap into place). A frame will appear on the display around the station number per
occupied slot.
Close the front cover.
Press the button 2x. "Clean only" will appear on the display.
Push the
(Program Start) button on the control membrane keyboard.
Program runs automatically. During the cleaning program, a clock will count backwards until 00:00 on the
display