Ensure that cassette is clean and has been treated with Stat Dri. Drain tube must run
directly to condenser bottle with no dips or loops. Do not stack wraps. Invert mesh
rack to provide air space below wraps. Set bubble level to 4 or 5 o'clock position.
Cover Removal
Unplug Statim and remove cassette, filler cap, 4 side screws and 3 rear screws.
On the Statim 5000 the middle screw in the rear is behind the biological filter holding the filter
bracket.
With facia overhanging edge of counter or table, lift up on rear of cover until it
clears internal components, then slide cover forward about an inch.
Gently rotate the cover up and pivot it around the front left corner of the machine.
When the cover is clear of the machine, rest it on its side. Take care not to strain
the ribbon cables connecting the display and keypad.
Refer to illustration on next page
Solenoid Valve Inspection and Repair
(Honeywell solenoid only)
Ensure that the power is off. Disassembly of the solenoid valve with power on
will cause valve coil damage.
Remove retaining nut, yoke and coil from top of valve.
Unscrew the plunger tube assembly. The tube may be removed by carefully