STATIM
onto the check valve inlet and pull the tube off the valve.
?. Disconnect the black wire from the PCB connector terminal J1-3. This black
wire is integral to the thermal fuse assembly.
3.Trace the path of the black wire back to the base of the steam generator (boiler).
Cut all the cable ties holding the black wire.
4.Disconnect the white wire attached to the terminal on the lower half of the steam
generator (boiler).
5. Using a 3/8 inch wrench, disconnect the compression nut holding the white
Teflon tube from the top of the steam generator (boiler).
6. Using a 7/16 wrench, disconnect the compression nut holding the steam Venerator
(boiler) outlet tube to the probe bracket inlet fitting.
7. Using a 7/16 wrench disconnect the compression nut holding the steam
generator (boiler) outlet tube to the steam generator (boiler) outlet fitting.
8. Remove the two screws from the steam generator (boiler) bracket. The steam
generator (boiler) is still attached to the PCB by the thermocouple lead.
9. Gently lift and turn the steam generator (boiler) assembly onto one side to
expose the bottom of the assembly. BE CAREFUL NOT TO STRESS THE
THERMOCOUPLE LEADS.
10.Remove the screw, which attaches the steam generator (boiler) bracket to the steam
generator (boiler). This screw is shorter than the screws used to hold the bracket to the
chassis.
11. Disconnect the thermal fuse thermal from the lower power terminal on the steam
generator (boiler). Note the routing of the thermal fuse assembly between the bracket and
spacing standoffs on the bottom of the steam generator (boiler).
http://www.sterilizers.com/statim-trouble-shooting-manual.html (13 of 26) [3/27/2004 9:23:44 PM]