STATIM
inserted you should hear the solenoid valve click.
5. Once installed and functioning, a thin bead of RTV silicone sealant may be applied along
the edges of the microswitch if available.
Reclaiming Statim Boiler
DO NOT RECLAIM BOILERS ON STATIMS THAT CONTAIN A
WATER QUALITY SENSOR.
I. Remove the pressure relief valve (PRV) from the top of the boiler.
2. Pour 1 oz. of CLR into the boiler through the PRV hole.
3. Reinstall the PRV using Teflon tape.
4. With the cassette inserted start an unwrapped cycle. As soon as you hear the pump start to
run press STOP. Allow the solution to work for 15 minutes.
5. Check that the reservoir is filled with good distilled water and the condenser bottle is not
full.
6. After 15 minutes have passed, start an unwrapped cycle while
simultaneously activating the pump.
You may operate the pump by shorting pins #16 & #18 on the 20 pin
connector (P I) on the PCB.
As long as the pins are shorted the pump will run continuously and flush the
remaining deposits out of
the boiler. Allow the pump to run for 3 minutes minimum.
7. Press STOP then start a normal unwrapped cycle.
IMPORTANT
8. You may verify that the boiler contamination problem has been corrected by
confirming that the pressure information (xxxKpA) appears on the display within ten
seconds after the beginning of the sterilization phase of a cycle.