SciCan STATIM Troubleshooting Manual Download Page 15

STATIM

10. Route the black wire from the fuse assembly to the PCB connector J1-3 and connect the wire. 

 

11. Bundle the loose wires together, using nylon cable ties every 2-3 inches. 

12. Carefully push the compressor tube onto the check valve inlet and secure 
the tube to the valve using a high temperature application cable tie. 

Microswitch Removal and Replacement

 

'To remove the microswitch follow these steps: 

1. Turn the main power switch OFF, and unplug the unit.

 

2. Disconnect the microswitch leads from the PCB block terminal positions labeled 
CASSIN and carefully cut the cable ties securing the microswitch leads. 

 

3.There may be RTV silicone sealant on the heads of the two screws securing the 
microswitch to the side of the probe bracket. Trim away the excess RTV silicone and 
remove the screws. Do Not remove the probe bracket to get at the screws. If you have 
trouble getting at the screws disconnect the gray hose between the check valve and the 
compressor and remove the four compressor mounting screws. Set the compressor off the 
back of the Statim being careful not to damage the wiring. 

 

4. Remove the microswitch. 

5. Remove all RTV silicone residue from the probe bracket surface. 

To replace the microswitch follow these steps: 

1. Attach the replacement microswitch to the probe bracket.

 

2. 

Connect the microswitch leads to the PCB block terminal positions labeled CASSIN. 

Secure the stripped leads. 

 

3. Route the microswitch leads the same as before and secure with nylon cable ties. 

 

4. 

Plug in the power cord and turn on the main power switch. When the cassette is fully 

http://www.sterilizers.com/statim-trouble-shooting-manual.html (15 of 26) [3/27/2004 9:23:44 PM]

Summary of Contents for STATIM

Page 1: ...f these conditions occur check your cassette and seal All leaks should be corrected before proceeding with troubleshooting procedures Check Cassette 1 Chamber temperature does not rise more than 2 C i...

Page 2: ......

Page 3: ......

Page 4: ......

Page 5: ...Calibration Section Check for kinks in the exhaust tubing Check to see if condenser bottle is full of hot water Cycle Fault 9 N A Cycle Fault 10 The cassette temperature has failed to drop to 115 C d...

Page 6: ......

Page 7: ......

Page 8: ...open and close and the solenoid clicks Repair cassette leak and clicking should stop in approximately 24 hours Loud buzzing noise Clean or replace solenoid as needed Noise during drying cycle only So...

Page 9: ...of cover until it clears internal components then slide cover forward about an inch Gently rotate the cover up and pivot it around the front left corner of the machine When the cover is clear of the m...

Page 10: ......

Page 11: ...holes in the new seal with the holes in the lid 6 Install the holes first then install the four corners allowing the sides to hang out Be sure that two indicator squares are visible in each corner 7 S...

Page 12: ...imately 800 cycles and if maintained properly could last longer Pump Tester Instructions I Disconnect the Teflon boiler inlet tube from the top of the boiler 2 Connect the disconnected end of the inle...

Page 13: ...he compression nut holding the steam generator boiler outlet tube to the steam generator boiler outlet fitting 8 Remove the two screws from the steam generator boiler bracket The steam generator boile...

Page 14: ...ting Tighten finger tight 5 Connect the compression nut holding the steam generator boiler outlet tube to the probe bracket inlet fitting Tighten finger tight 6 Connect the compression nut holding the...

Page 15: ...e bracket Trim away the excess RTV silicone and remove the screws Do Not remove the probe bracket to get at the screws If you have trouble getting at the screws disconnect the gray hose between the ch...

Page 16: ...the reservoir is filled with good distilled water and the condenser bottle is not full 6 After 15 minutes have passed start an unwrapped cycle while simultaneously activating the pump You may operate...

Page 17: ...d chamber temperatures The value in the lower right corner of the display is the chamber temperature F9 CALIBRATING C8 2 00 C 135 00 C EXAMPLE 5 To start the steam generator boiler self calibration cy...

Page 18: ...f the Statim on Press the WRAPPED button again to scroll to the next language Press the UNWRAPPED button to scroll to the previous language When the desired language is displayed press the STOP button...

Page 19: ...button Press the button repeatedly to select subsequent fields 7 To save changes and return to regular operating mode press the STOP button Diagnosing Water Quality Sensor Problems The water quality...

Page 20: ......

Page 21: ......

Page 22: ......

Page 23: ......

Page 24: ......

Page 25: ...x 7 1 2 Cassette Internal Dimensions 15 x 7 x 1 Reservoir Capacity 4 liters distilled water Approximately 20 cycles Weight Without Water 72 lbs Power Consumption 1300 W Optional Printer Type Thermal...

Page 26: ...ber and Plastics 40 minutes Heavy Duty Unwrapped 12 minutes Warranty I Year Parts excluding gasket 1 Year SciCan Labor http www sterilizers com statim trouble shooting manual html 26 of 26 3 27 2004 9...