STATIM
900J Operator’s Manual
6
3.1 Environmental Considerations
There are several factors that may effect the performance of your
Statim. Please review these factors, and select a suitable location in
which to install the unit.
• Temperature and Humidity
Avoid installing your Statim in direct sunlight or close to a heat source
(e.g. vents or radiators). The recommended operating temperature is
between 15 ˚C to 25 ˚C (59 ˚F to 77 ˚F) with humidity of 25 % to 70 %.
• Spacing
The vents and openings on the Statim should remain uncovered and
unobstructed. Leave a minimum of 50 mm / 2 " between the top, sides
and back of the unit and any wall or partition.
• Venting
The Statim should be operated in a clean, dust-free environment.
• Work Surface
The Statim should be placed on a flat, level, water-resistant surface.
Never install and operate the unit on a sloped surface.
• Electromagnetic Environment
The Statim has been tested and meets applicable standards
for electromagnetic emissions. While the unit does not emit any
radiation, it may itself be affected by other equipment which does.
We recommend that the unit be kept away from potential sources of
interference.
• Electrical Requirements
Use properly grounded and fused power sources with the same voltage
rating as indicated on the label at the back of your Statim. Avoid
multiple outlet receptacles. If using a surge suppressor power bar, plug
in one Statim only.
3.2 Unit Placement
To stabilize the Statim 900J, follow these steps:
1. Align the bubble (1)
in the level indicator
(2) on the front panel
to the right front quadrant
(or 4 o’clock) of the target
by adjusting the four
leveller feet (3). This will
ensure that the Statim 900
functions properly.
2. Ensure that all four feet are
in contact with the surface,
and that the Statim 900J
does not rock.
1. Bubble
2. Level indicator
3. Leveller feet
4. Reservoir
cap
Figure 3
4
2
1
3 (4x)
3. Installation