STATIM
900J Operator’s Manual
15
2. Hollow or Wrapped Instrument Cycle
The Hollow or Wrapped Instrument Cycle is used to sterilize dental handpieces.
The maximum recommended load for the Handpiece cycle is 1.0 kg.
To select this cycle, press the Hollow or Wrapped Instrument Cycle button.
The cycle then starts automatically. This cycle is illustrated in the graph below.
During the sterilizing phase of this cycle, the temperature in the chamber is
maintained at a minimum of 134ºC for 5 minutes. See Preparing and Loading
Instruments before running this cycle.
3. Rubber or Plastic Instrument Cycle
The Rubber or Plastic Instrument Cycle is used to sterilize instruments
constructed of the materials listed under the heading Rubber or Plastic
Instruments in Preparing and Loading Instruments. The maximum recommended
load for the Rubber or Plastic Instrument Cycle is 0.1 kg.
To select this cycle, press the Rubber or Plastic Instrument Cycle button. The
cycle then starts automatically. This cycle is illustrated in the graph below.
During the sterilizing phase of this cycle, the temperature in the chamber is
maintained at a minimum of 121ºC for 15 minutes. See Preparing and Loading
Instruments before running this cycle.
4. Instructions for Use
HOLLOW OR WRAPPED 134°C / 5 min
A
B
C
E
D
A
Warming up
B
Conditioning
C
Pressurizing
D
Sterilizing
E
Venting
RUBBER / PLASTIC 121°C / 15 min
A
B
C
E
D
A
Warming up
B
Conditioning
C
Pressurizing
D
Sterilizing
E
Venting