STATIM 7000 Service Manual –
Validation Solenoid Valve
26
7. Remove the 2 screws that attach
the validation solenoid valve
bracket to the chassis.
8. Remove 2 screws on the under side
of the unit that hold solenoid housing
to the chassis.
9. Remove solenoid.
10. Reverse removal instructions to