STATIM 7000 Service Manual – Keypad & LCD
8
4. Membrane Keypad and LCD
The membrane keypad and LCD contain circuitry that is static sensitive.
Always wear a static strap when working with or near this circuitry. Transport
electronic components in a static protected bag.
To replace the membrane keypad or LCD, perform the following steps:
1. With the unit off, unplug the power cord from the wall outlet and remove
the cassette and reservoir from the unit.
2. Remove the water filter or water bypass cartridge (if the unit is using
distilled water) from the reservoir area.
3. Remove the eight screws across the bottom front of the unit.
4. Push the cover forward from the back to loosen it.
5. Detach the LCD by reaching up
inside the cover through the
armature opening to feel for a
plastic tab located directly
behind the Rubber and Plastics
key.
Push this tab to the left to unlock the
LCD from the cover.