7. Steam Generator
To remove the steam generator, follow these steps:
1.
Power
OFF
the unit, unplug it and remove the cover. See Cover. NOTE:
The steam generator may be hot if the unit has been operating.
2.
Disconnect the fan connectors.
3.
Disconnect the inlet and outlet tubes on the biological air filter.
4.
Remove the rear chassis cover plate by removing the three screws across
the top and three screws across the bottom, and lifting the cover plate.
5.
Disconnect the steam generator thermocouple
from the controller board.
6.
Cut the tie wraps holding the steam generator thermocouple.
7.
Disconnect black steam generator thermal fuse wire from controller board
connector terminal block.
8.
Disconnect the white steam
generator thermal fuse wire from
the upper terminal in the steam