STATIM 7000 Service Manual – Steam Gen. Thermal Fuse
18
9. Steam Generator Thermal Fuse
To remove the steam generator thermal fuse, follow these steps:
1. Power
OFF
the unit, unplug it and remove the cover. See Cover. NOTE:
The steam generator may be hot if the unit has been operating.
2. Disconnect the fan connectors.
3. Disconnect the inlet and outlet tubes on the biological air filter.
4. Remove the rear chassis cover plate by removing the 3 screws across the
top and 3 screws across the bottom, and lifting the cover plate.
5. Disconnect the black thermal fuse lead wire from Controller Board
connector terminal and the white wire from the steam generator.
6. Trace the path of the black wire back to the base of the steam generator.
Cut all the cable ties holding the black wire.
7. Disconnect the compression nut
holding the Teflon™ tube from
the top of the steam generator.