3
Validation Thermocouple Calibration:
11.
New!
Turn the machine off and back on, while keeping Unwrapped and Wrapped
button pressed to go back to
Service Mode
. Enter password to continue. Toggle
using keypad through the menu selection to reach Calibration option and press R&P
key. The display should show the calibration screen.
12. Check that there is sufficient water in the water reservoir before proceeding.
13. Start a Validation Assy. self-calibration cycle. To do this, press and
hold
the
UNWRAPPED button and in the same time press START button. The character “
∗
”
will appear immediately to the right of the Validation thermocouple hexadecimal
offset on the display to indicate that a Validation Assy. calibration cycle is running.
This calibration will take approximately 6 minutes.
14. Allow the Validation Assy. self-calibration to complete. The temperature within the
chamber will rise to the sterilization temperature. Wait until sterilization phase of the
calibration cycle ends automatically. The offset value in the upper left-hand corner
of the display (CVO) may have changed to a new offset value.
15. Press the STOP button to end the Validation TC self-calibration cycle.
16. Power off the Statim.
Calibration of the Statim autoclave thermocouples is now complete.