10. Statim 5000 Trouble-shooting – Printer Messages
47
Printer Fault
Description of Fault
Suggested steps for Correction
of fault
Message: PRINTER
FAULT (if optional printer is
installed)
Printer is not printing
This is caused by either a paper
jam or a defective printer controller
board.
"CYCLE ABORTED"
This error message is displayed on the printout only, followed by
the message "NOT STERILE", as a result of the operator
pressing the STOP button to stop the cycle or as a result of any
other abnormal cycle termination, including CYCLE FAULT
errors.
"STOP BUTTON
PRESSED"
The operator pressed the STOP button to stop the cycle. The
LCD shows the message "NOT STERILE" as a result.
"CYCLE
INTERRUPTED"
This error is displayed on power-up following a power failure
occurs during a cycle or whenever the power is turned OFF
after an error occurred without pressing the STOP button to
reset. (Rev. R200 level code)
This message is displayed when the sterilization phase has
failed to start within three minutes of the cassette reaching the
sterilization temperature. If it occurs in three consecutive cycles
Cycle Fault #26 is displayed. (Software Rev. R4xx and R5xx)
"PRESS STOP TO
RESET"
This message is displayed for all error faults. The user MUST
press the STOP button on the keypad to reset the unit;
otherwise the user will be unable to initiate another cycle.
No message displayed
The printer does not work
Make sure that all printer cables
are connected. Check connections
at the Controller Board and the
Printer Controller Board. The cable