21
2. Service Menu
Note: To enter Service Menu power up Statim while holding Unwrapped and Wrapped
keys pressed. After that enter password to access Service menu. Default password is
Unwrapped, Wrapped, R&P and Stop keys in this order. The service technician can
change this password. In case the changed password is lost a backdoor password can
be used: Unwrapped, Wrapped, Unwrapped, Wrapped in this order.
To setup the advanced options of Statim
>Calibration
Time/Date Setup
Language Setup
Unit ID Setup
Set cycle counter
Lock cycles
For S class units only
Conductivity Setup
Water.Cnd Tmp. Comp
Last Printout
Stored CF Printouts
Clear CF Printouts
Display last CF#
Devices Test On/Off
Temperature Offset
Press. Atm. Offset
For S class units only
Press. Ster. Offset
For S class units only
Press. Sensor Type
For S class units only
Validation Offset
For non-S class units only
Voltage Calibration
Voltage setup
Repeater mode
RS232
End of Line CR/LF
Serial Port Bitrate
Printer user ß char
Factory default
Change Password
Backup NVRAM
Restore NVRAM
Save and Exit
Exit
Production Cycle
Keypad
Unwrapped
Select next item in the menu
Wrapped
Select previous item in the menu
Rubber and Plastic
Enter the indicated sub menu selection
Stop
Exit and return to normal mode of operation