24
To start the Chamber thermocouple Calibration cycle, first
press and release the Unwrapped key followed by pressing
the Start key.
Wrapped
Sends the last printout to the printer. The printer must be
connected and selected for this feature to work.
Rubber and Plastics
Start the Burn-in cycle
Stop
Exit to normal mode of operation
Keypad functions when calibration cycle is running:
Unwrapped
Increment current selected offset.
Wrapped
Decrement current selected offset
Rubber and Plastics
Select other offset
Stop
Interrupt calibration cycle
2.2. Set cycle counter
Adjust the recorded number of cycles ran
Cycle #
000000
Keypad
Unwrapped
Decrement current digit
Wrapped
Increment current digit
Rubber and Plastics
Move to next digit
Stop
Save and exit to normal mode of operation
2.3. Lock cycles
Expands or contracts the selection of cycle parameters for each cycle button.
Keypad
Unwrapped
Toggle to Locked/Unlocked mode
Stop
Save and exit to normal mode of operation
Display after toggling to Unlocked mode
Cycles UNLOCKED
Available cycles
Display after toggling to Locked mode
Cycles LOCKED
Selected cycles
In the “Cycles LOCKED” mode, each cycle button will represent one set of cycle
parameters. In the “Cycles UNLOCKED” mode, each cycle button can represent
multiple sets of cycle parameters. After toggling to each mode, the corresponding