Treatment factor (over/ undercorrection):
If the clinical study shows in some parts over or undercorrections the user has to
change here the treatment factors for spherical part for myopia and hyperopia,
for the myopic astigmatism(neg.) and the hyperopic astigmatism (pos.) the cylindrical
value.
Myopia, Hyperopia, myopic Astigmatism and hyperopic
Astigmatism:
On the right side the user can limitate the maximum refraction values for myopia,
hyperopia, myopic astigmatism and hyperopic astigmatism.
Laser:
On the bottom of the formular the service can store the laser energy, which is
responsible for all treatment methodes.
The laser frequency is only displayed, but not changeable.
The number of test-pulses has the meaning that the service can increase or
decrease the number of laser pulses during the necessary tests on photo paper, like
scanner test, scanner drift, concentic test and so on. Sometimes, if is it necessary,
because the service is using different types of photopaper with different thickness.
Distributor adress
By selecting the menu distributor adress the service can store the adress of the
manufacturer
Schwind
or the
distributor
. This is interesting to show the user, who
is responsible for gas orders or service cases. Everytime, when something is
happening or necessary the software will display the adress on the screen.