7 Installation of the ESIRIS-Laser
Assembling the Laser
1.
Move the laser rack to the correct place in the room. Then screw down the 4 feet
of the rack until the wheels rotate free.
2. Unscrew all 4 wheels and put them away.
3. Screw down all the 4 feet step by step, not one after the other, otherwise the
laser could cant (Danger!!). Screw the feet down to a distance of 8 cm from floor
to the bottom edge of the laser rack.
4. Align the laser rack with a spirit level by minimally moving the feet.
Assembling of the Patient Bed
1. Disconnect the light fibers for lasertrigger and to the RS 232 interface.
2. Move the patient bed carefully to the laserrack until both fit together. Be careful
not to snap off the gas pipe!