schwa-medico
Instructions for use - AS SUPER 4 digital
24
2.
Warning:
Keep the product away from water and other liquids
as this may cause uncontrolled current flows and electric shocks and damage to the
product.
3.
Caution:
Consult a physician before
applying electrodes over or through
the head, directly on the eyes, covering the mouth, on the front of the neck, (espe-
cially the carotid sinus), or from electrodes placed on the chest and the upper back or
crossing over the heart.
Warning:
Application of electrodes near the thorax may increase the risk of car-
diac fibrillation. In case of electrode placement in the thorax area, intensive high
frequency stimulation (above approx. 15 Hz) may lead to respiratory disorders during
stimulation.
4.
Warning:
It is not allowed to use the product during operating machines or during
operations which needs elevated concentration. Pay attention particularly at road
traffic.
5.
The product may only be used with original accessories.
The use of other accessories (especially electrodes with a smaller surface than 2 cm²)
may lead to a deficient operation.
6.
Caution:
Simultaneous connection of a patient to a high frequency surgical ME
(medical electrical) equipment may lead to burns beneath the electrodes and it may
damage to the product.
7.
Caution:
Operation in the vicinity (e.g. 1m) of a short wave or microwave ME de-
vice for shortwave or microwave therapy may cause fluctuations in the output value
of the product which may turn into painful reactions.
8.
Caution:
Keep a distance of min. 30 cm (12 inches) between the parts and the
wires of this product and a high frequency communication device (mobile phone or a
radio device including their accessories such as antenna cable or external antenna).
Non-observance may result in a reduction of the device’s performance characteristics
and a deficient operation.
9.
Caution:
Do not use this device next to or stacked with other devices, as this may
result in deficient operation. If, however, use in the manner described above is still
necessary, this device and the other devices should be observed to ensure that they
are working properly.
10.
Device mains plugs and/or socket outlet must be easily accessible
during operation of the device,
so that disconnection from the mains supply
is quickly possible if necessary.
11.
Placing of the needles/electrodes:
a. The device may only be connected to just 1 patient.
b. Before applying the needles/electrodes, clean the skin surface on which the elec-
trodes are to be attached. Otherwise incorrect operation cannot be excluded.
c. Take care that no metallic objects (e.g. jewellery or piercings) get in contact with
the needles/electrodes during the stimulation because otherwise isolated skin
burns may appear.