3
English
Intended use
The SPORÉCUP XTR2 has been designed as a device for the generation of
transcutaneous electrical stimulation of the nerves and muscles on humans and
may not be used for any other purposes. Please read the operating instructions
carefully before using the SPORÉCUP XTR2.
Safety Instructions
To ensure safe usage of the product
Please read the instructions for use carefully prior to using the product!
1
The XTR2 may only be used with original accessories.
2
Keep XTR2 away from water and other liquids.
3
Never use the XTR2 if it functions incorrectly or has been damaged.
4
Advice that stimulation should not be applied across or through the head,
directly on the eyes, covering the mouth, on the front of the neck, (especially
the carotid sinus), or from electrodes placed on the chest and the upper back
or crossing over the heart.
5
The used electrodes must not be smaller than 2 cm².
6
Any electrodes that have current densities exceeding 2 mA/cm² may require
the special attention of the operator.
7
It is only allowed to connect up the XTR2 to one person.
8
It is not allowed to use the XTR2 during operating machines or during opera-
tions which needs elevated concentration. Pay attention particulary at road
traffic.
9
Do not allow XTR2 to be dropped, handled incorrectly, exposed to extreme
temperatures or high humidity (only use at temperatures between 10 °C and
40 °C and a relative humidity of less than 90 %).
10
Care must be taken when the XTR2 is used on or in the proximity of children.
11
Store the XTR2 in the original packaging after use to protect it against damage
and soiling.
Warning!
»
The simultaneous connection of the patient to a high-frequency surgical device
may lead to burns beneath the electrodes. Operation in the vicinity of a short
wave or microwave device may cause fluctuations in the output value of the
XTR2.
Summary of Contents for SPORECUP XTR2
Page 23: ...23 English ...