Simultaneous connection of a patient to a high frequency surgical ME equipment
may result in burns at the site of the stimulator electrodes and possible damage
to the XTR2.
»
Operation in close proximity (e.g. 1 m) to a shortwave or microwave therapy ME
equipment may produce instability in the XTR2 output.
»
Application of electrodes near the thorax may increase the risk of cardiac
fibrillation.
Contraindications
In which cases shall a patient not use the device – or if – only with consultation
of a medical practitioner?
»
Patients with implanted electronic devices (pace maker etc.)
»
Patients with cardiac dysrhythmia
»
Patients with Seizure disorder (epilepsy)
»
Patients with skin irritations in the area of the electrode application
»
Patients with malignant diseases (tumours) in the area of the application
Contraindications TENS/EMS in pregnancy
In addition to the general contraindications of TENS/EMS applies:
»
The use of TENS during pregnancy should always be coordinated with the atten-
ding physician and the midwife, taking into account the benefits and the risks
»
Patients who have experienced abortion or premature delivery, should not use
TENS during pregnancy
»
Patients with preterm labour TENS should not use TENS during pregnancy
»
TENS should not be used during the first three months of pregnancy or only
after a careful risk assessment. In particular, stimulation in the area of the uterus
should be avoided
»
From the 4th month of pregnancy onwards TENS should not be applied near
the uterus. This applies to all electrode placements in the area of the abdomen,
pelvis and lower back
»
The use of TENS is permissible during childbirth
Side effects
»
Pain caused by the stimulation: The stimulation can be perceived as uncomfor-
table/painful if the electric current is too intensive or if the electrodes have been
positioned unfavorably. Unpleasant side effects can be avoided by adjusting
the intensity, using another program with other parameters and the possible
re-positioning of the electrodes.